Feasibility of patient‐specific quality assurance (PSQA) for real‐time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14352 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Qianyi Xu und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Advanced Radiation Oncology and Proton Therapy, Inova Schar Cancer Institute, Fairfax, Virginia, USA. Hauptthemen laut MEDLINE: Radiosurgery, Quality Assurance, Health Care, Phantoms, Imaging, Feasibility Studies. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To design a patient specific quality assurance (PSQA) process for the CyberKnife Synchrony system and quantify its dosimetric accuracy using a motion platform driven by patient tumor traces with rotation.
Methods
The CyberKnife Synchrony system was evaluated using a motion platform (MODUSQA) and a SRS MapCHECK phantom. The platform was programed to move in the superior‐inferior (SI) direction based on tumor traces. The detector array housed by the StereoPhan was placed on the platform. Extra rotational angles in pitch (head down, 4.0° ± 0.15° or 1.2° ± 0.1°) were added to the moving phantom to examine robot capability of angle correction during delivery. A total of 15 Synchrony patients were performed SBRT PSQA on the moving phantom. All the results were benchmarked by the PSQA results based on static phantom.
Results
For smaller pitch angles, the mean gamma passing rates were 99.75% ± 0.87%, 98.63% ± 2.05%, and 93.11% ± 5.52%, for 3%/1 mm, 2%/1 mm, and 1%/1 mm, respectively. Large discrepancy in the passing rates was observed for different pitch angles due to limited angle correction by the robot. For larger pitch angles, the corresponding mean passing rates were dropped to 93.00% ± 10.91%, 88.05% ± 14.93%, and 80.38% ± 17.40%. When comparing with the static phantom, no significant statistic difference was observed for smaller pitch angles ( p = 0.1 for 3%/1 mm), whereas a larger statistic difference was observed for larger pitch angles ( p < 0.02 for all criteria). All the gamma passing rates were improved, if applying shift and rotation correction.
Conclusions
The significance of this work is that it is the first study to benchmark PSQA for the CyberKnife Synchrony system using realistically moving phantoms with rotation. With reasonable delivery time, we found it may be feasible to perform PSQA for Synchrony patients with a realistic breathing pattern.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Advanced Radiation Oncology and Proton Therapy, Inova Schar Cancer Institute, Fairfax, Virginia, USA.
- Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania, USA.
- Department of Radiation Oncology, University of Miami, Miami, Florida, USA.
- Department of Radiation Oncology, Taizhou Hospital, Taizhou, Zhejiang, China.
- Department of Radiation Oncology, Wellstar Health System, Marietta, Georgia, USA.
ORCID-Profile: Haihua Yang
Zitieren
APA
Xu, Q., Fan, J., Vinogradskiy, Y., Chawla, A. K., Kubicek, G., Yang, H., Huynh, K., LaCouture, T., Grimm, J., & Nie, W. (2024). Feasibility of patient‐specific quality assurance (PSQA) for real‐time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14352
Vancouver
Xu Q, Fan J, Vinogradskiy Y, Chawla AK, Kubicek G, Yang H, et al. Feasibility of patient‐specific quality assurance (PSQA) for real‐time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14352
BibTeX
@article{Xu2024,
author = {Xu, Qianyi and Fan, Jiajin and Vinogradskiy, Yevgeniy and Chawla, Ashish K and Kubicek, Gregory and Yang, Haihua and Huynh, Kiet and LaCouture, Tamara and Grimm, Jimm and Nie, Wei},
title = {Feasibility of patient‐specific quality assurance (PSQA) for real‐time robotic stereotactic body radiotherapy (SBRT) based on tumor motion traces},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14352}
}
EUVIMED