The delivery assessment for small targets on Halcyon radiotherapy system – Measured and calculated dose comparison

Linda Lankinen iD, Antti Kulmala, Jouko Lehtomäki, Ari Harju

Journal of Applied Clinical Medical Physics · 2024-06 · 25 (6)

DOI: 10.1002/acm2.14407 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 22.05.2024. Verfasst von Linda Lankinen und 3 weiteren Autor:innen. Beteiligte Einrichtung: Varian Medical Systems, a Siemens Healthineers Company, Helsinki, Finland. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage, Radiotherapy, Intensity-Modulated, Monte Carlo Method. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Varian Medical Systems, A Siemens Healthineers Company.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

With the ever‐increasing requirements of accuracy and personalization of radiotherapy treatments, stereotactic radiotherapy (SRT) with volumetric modulated arc therapy (VMAT) on O‐ring Halcyon radiotherapy system could potentially provide a fast, safe, and feasible treatment option.

Purpose

The purpose of this study was to assess the delivery of Halcyon VMAT plans for small targets.

Methods

Well‐defined VMAT‐SRT plans were created on Halcyon radiotherapy system with the stacked and staggered dual‐layer MLC design for the film measurement set‐up and the target sizes and shapes designed to emulate the targets of the stereotactic treatments. The planar dose distributions were acquired with film measurements and compared to a current clinical reference dose calculation with AcurosXB (v18.0, Varian Medical Systems) and to Monte Carlo simulations. With the collapsed arc versions of the VMAT‐SRT plans, the uncertainty in dose delivery due to the multileaf collimator (MLC) without the gantry rotation could be separated and analyzed.

Results

The target size was mainly limited by the resolution originated from the design of the MLC leaves. The results of the collapsed arc versions of the plans show good consistency among measured, calculated, and simulated dose distributions. With the full VMAT plans, the agreement between calculated and simulated dose distributions was consistent with the collapsed arc versions. The measured dose distribution agreed with the calculated and simulated dose distributions within the target regions, but considerable local differences were observed in the margins of the target. The largest differences located in the steep gradient regions presumably originating from the deviation of the isocenter.

Conclusions

The potential of the Halcyon radiotherapy system for VMAT‐SRT delivery was evaluated and the study revealed valuable insights on the machine characteristics with the delivery.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-ModulatedMonte Carlo MethodOrgans at RiskHumansPhantoms, ImagingParticle AcceleratorsNeoplasmsRadiosurgery

Stichwörter der Autor:innen

HalcyonMonte CarloVMATsmall fieldstereotactic

Autor:innen und Institutionen

  • Varian Medical Systems, a Siemens Healthineers Company, Helsinki, Finland.
  • Department of Physics, University of Helsinki, Helsinki, Finland.
  • Clinical Research Institute HUCH Ltd., Helsinki, Finland.

ORCID-Profile: Linda Lankinen

Förderung

  • Varian Medical Systems
  • A Siemens Healthineers Company

Zitieren

APA

Lankinen, L., Kulmala, A., Lehtomäki, J., & Harju, A. (2024). The delivery assessment for small targets on Halcyon radiotherapy system – Measured and calculated dose comparison. Journal of Applied Clinical Medical Physics, 25(6). https://doi.org/10.1002/acm2.14407

Vancouver

Lankinen L, Kulmala A, Lehtomäki J, Harju A. The delivery assessment for small targets on Halcyon radiotherapy system – Measured and calculated dose comparison. Journal of Applied Clinical Medical Physics. 2024;25(6). doi:10.1002/acm2.14407

BibTeX
@article{Lankinen2024,
  author = {Lankinen, Linda and Kulmala, Antti and Lehtomäki, Jouko and Harju, Ari},
  title = {The delivery assessment for small targets on Halcyon radiotherapy system – Measured and calculated dose comparison},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {6},
  doi = {10.1002/acm2.14407}
}
Kostenlos starten →