Commissioning and verification of a 3D Monte Carlo independent calculation software for O‐ring linac systems
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70445 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Xiangyin Meng und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Peking Union Medical College Hospital, Beijing, China. Hauptthemen laut MEDLINE: Monte Carlo Method, Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Particle Accelerators. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Key R&D Program of China, Ministry of Science and Technology of the People's Republic of China, National High Level Hospital Clinical Research Funding.
Abstract (Originalsprache)
Background
O‐ring linac systems improve radiotherapy efficiency but require rigorous pretreatment verification due to increased delivery uncertainty in IMRT/VMAT. Existing methods face limitations: measurement‐based approaches incur setup errors, while calculation‐based methods (e.g., Monte Carlo) need machine‐specific validation per AAPM TG‐219.
Purpose
To commission the O‐ring linac model in RadCalc Monte Carlo software and establish its clinical dosimetric accuracy.
Methods
The additional radiation to light field offset (ARLFO) parameter in RadCalc was adjusted from −0.08 cm (−0.8 mm) to +0.08 cm (+0.8 mm) (nine sets). TG‐119 and clinical benchmark cases were used to design IMRT/VMAT plans. Verification plans were measured experimentally, and all plans were imported into RadCalc for secondary dose calculation. Triangular gamma analysis (3%/2 mm) compared Monte Carlo‐simulated, measured, and TPS‐calculated doses. The optimal ARLFO was determined by weighted averaging of gamma pass rates. The model was validated on 10 clinical cases per site (head, thorax, abdomen, pelvis).
Results
Commissioning identified the optimal ARLFO parameter as −0.02 cm (−0.02 mm). Under the gamma analysis criterion of 3%/2 mm, the comparison between TPS‐calculated doses and Monte Carlo‐calculated doses for 10 clinical plans across four anatomical sites yielded: 98.9% ± 0.8% (head and neck), 98.5% ± 0.8% (thorax), 99.6% ± 0.3% (abdomen), and 99.1% ± 0.5% (pelvis). For verification plans, the gamma pass rates between Monte Carlo‐calculated doses and measured doses were 97.5% ± 2.8% (head and neck), 95.6% ± 2.8% (thorax), 96.3% ± 2.9% (abdomen), and 99.4% ± 0.5% (pelvis), while comparisons of Monte Carlo‐calculated doses versus TPS‐calculated doses reached 99.5% ± 0.8% (head and neck), 99.2% ± 1.0% (thorax), 99.5% ± 0.9% (abdomen), and 99.9% ± 0.2% (pelvis), demonstrating consistent dosimetric accuracy of the optimized model across all clinical sites.
Conclusion
This study establishes a commissioning methodology to determine the optimal ARLFO value for RadCalc, enabling clinics to achieve reliable independent plan verification for O‐ring linac.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Peking Union Medical College Hospital, Beijing, China.
- LAP Laser Applications China Co., Ltd., Shanghai, China.
ORCID-Profile: Xiangyin Meng, Zhen Zhou, Zhiqun Wang
Förderung
- National Key R&D Program of China
- Ministry of Science and Technology of the People's Republic of China
- National High Level Hospital Clinical Research Funding
- National Key Research and Development Program of China
Zitieren
APA
Meng, X., Dong, T., Liang, Y., Zhou, Z., Wang, Z., Qiu, J., Zhao, J., Zhang, Z., Li, W., & Yang, B. (2026). Commissioning and verification of a 3D Monte Carlo independent calculation software for O‐ring linac systems. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70445
Vancouver
Meng X, Dong T, Liang Y, Zhou Z, Wang Z, Qiu J, et al. Commissioning and verification of a 3D Monte Carlo independent calculation software for O‐ring linac systems. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70445
BibTeX
@article{Meng2026,
author = {Meng, Xiangyin and Dong, Tingting and Liang, Yongguang and Zhou, Zhen and Wang, Zhiqun and Qiu, Jie and Zhao, Jingjing and Zhang, Zheqing and Li, Wenbo and Yang, Bo},
title = {Commissioning and verification of a 3D Monte Carlo independent calculation software for O‐ring linac systems},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70445}
}