EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Practical considerations for handling implants in proton therapy

Gloria Vilches-Freixas iD, Vicki Trier Taasti iD, Ilaria Rinaldi iD, Mirko Unipan iD

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70322 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Gloria Vilches-Freixas und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology (Maastro), GROW Research Institute for Oncology and Reproduction, Maastricht University Medical Centre+, Maastricht,…. Hauptthemen laut MEDLINE: Proton Therapy, Radiotherapy Planning, Computer-Assisted, Tomography, X-Ray Computed, Prostheses and Implants. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

CT simulation scan is typically used for proton therapy planning. However, non‐tissue‐like materials are often not correctly characterized by the CT conversion curve, and medical implants of high‐Z materials, such as metals, can introduce artifacts in CT images. These can negatively affect the accuracy of dose calculations, ultimately impacting both treatment planning and dose delivery.

Purpose

To present our clinical procedures for managing implants in patients treated with proton therapy.

Methods

We categorized implants into two groups—‘small’ and ‘large’—based on their size relative to the CT voxel and their effect on image quality. To assess the influence of different implants on dose calculations, we conducted dose evaluations in a few patients with different implants. Our key considerations included the selection of appropriate CT window‐level settings for implant contouring, the necessity of performing material overrides weighing the tradeoff between the time spent on contouring and the impact on the dose distribution, as well as artifact mitigation strategies. The proposed procedures were derived from these dose evaluations and our clinical experience.

Results

For small implants, made of titanium or tantalum, we can rely on our CT conversion curves, without requiring material overrides. Conversely, larger implants, such as silicone breast implants, metal hip prostheses, or air pockets, require material overrides to ensure an accurate representation of stopping‐power ratios and to minimize dose discrepancies. The general rule is to avoid proton beams passing through metal implants when possible.

Conclusions

We have presented a comprehensive and standardized strategy for managing implants in our routine clinical practice. These strategies have been specifically developed for our clinical setup based on pragmatic considerations and may not be universally applicable. However, we believe they can offer valuable insights for other proton therapy centers facing similar challenges and looking to enhance their protocols for implant management.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Proton TherapyRadiotherapy Planning, Computer-AssistedTomography, X-Ray ComputedProstheses and ImplantsNeoplasmsHumansRadiotherapy DosageArtifactsRadiotherapy, Intensity-ModulatedPhantoms, ImagingOrgans at Risk

Stichwörter der Autor:innen

CT image artifactsclinical instructionsmetal implantsproton therapy

Autor:innen und Institutionen

  • Department of Radiation Oncology (Maastro), GROW Research Institute for Oncology and Reproduction, Maastricht University Medical Centre+, Maastricht, The Netherlands.
  • Danish Centre for Particle Therapy, Aarhus University Hospital, Aarhus, Denmark.

ORCID-Profile: Gloria Vilches-Freixas, Vicki Trier Taasti, Ilaria Rinaldi, Mirko Unipan

Zitieren

APA

Vilches-Freixas, G., Taasti, V. T., Rinaldi, I., & Unipan, M. (2025). Practical considerations for handling implants in proton therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70322

Vancouver

Vilches-Freixas G, Taasti VT, Rinaldi I, Unipan M. Practical considerations for handling implants in proton therapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70322

BibTeX
@article{VilchesFreixas2025,
  author = {Vilches-Freixas, Gloria and Taasti, Vicki Trier and Rinaldi, Ilaria and Unipan, Mirko},
  title = {Practical considerations for handling implants in proton therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70322}
}
Infokarte | EUVIMED