Fachartikel in der Zeitschrift Journal of Functional Biomaterials, online erschienen am 09.04.2024. Verfasst von Maryam Soleimani und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Engineering Materials and Biomaterials, Faculty of Mechanical Engineering, Silesian University of Technology, 18a Konarskiego Str., 41-…. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Re-tightening the loosened dental implant abutment screw is an accepted procedure, however the evidence that such screw will hold sufficiently is weak. The purpose of this study was material analysis of lost dental implant abutment screws made of the TiAlV alloy from various manufacturers, which became lost due to unscrewing or damaged when checking if unscrewed; undamaged screws could be safely re-tightened. Among 13 failed screws retrieved from 10 cases, 10 screws were removed due to untightening and 3 were broken but without mechanical damage at the threads. Advanced corrosion was found on nine screws after 2 years of working time on all surfaces, also not mechanically loaded. Sediments observed especially in the thread area did not affect the corrosion process because of no pit densification around sediments. Pitting corrosion visible in all long-used screws raises the question of whether the screws should be replaced after a certain period during service, even if they are well-tightened. This requires further research on the influence of the degree of corrosion on the loss of the load-bearing ability of the screw.

Schlagwörter

Stichwörter der Autor:innen

dental implant abutment screwfailurefracturelooseningpitting corrosiontitanium alloytitanium nitride coatingunscrewing

Autor:innen und Institutionen

  • Department of Engineering Materials and Biomaterials, Faculty of Mechanical Engineering, Silesian University of Technology, 18a Konarskiego Str., 41-100 Gliwice, Poland.
  • Doctoral School, Silesian University of Technology, 2A Akademicka Str., 44-100 Gliwice, Poland.
  • Anident Dental Clinic, 12 Belgradzka Str., 02-793 Warszawa, Poland.

ORCID-Profile: Maryam Soleimani, Jarosław Żmudzki, Wojciech Pakieła

Zitieren

APA

Soleimani, M., Żmudzki, J., Pakieła, W., Jaśkowska, A., & Krasny, K. (2024). Dental Implant Abutment Screw Loss: Presentation of 10 Cases. Journal of Functional Biomaterials, 15(4), 96. https://doi.org/10.3390/jfb15040096

Vancouver

Soleimani M, Żmudzki J, Pakieła W, Jaśkowska A, Krasny K. Dental Implant Abutment Screw Loss: Presentation of 10 Cases. Journal of Functional Biomaterials. 2024;15(4):96. doi:10.3390/jfb15040096

BibTeX
@article{Soleimani2024,
  author = {Soleimani, Maryam and Żmudzki, Jarosław and Pakieła, Wojciech and Jaśkowska, Anna and Krasny, Kornel},
  title = {Dental Implant Abutment Screw Loss: Presentation of 10 Cases},
  journal = {Journal of Functional Biomaterials},
  year = {2024},
  volume = {15},
  number = {4},
  pages = {96},
  doi = {10.3390/jfb15040096}
}
Kostenlos starten →