Effect of Abutment Screw Design on Torque Loss Under Cyclic Fatigue Loading: A Comparison of TSIII and KSIII Implant Systems

Jung-Tae Lee iD, Jae-Chang Lee, Dong-Wook Han iD, Bongju Kim iD

Bioengineering · 21.10.2025 · 12 (10) · 1131

DOI: 10.3390/bioengineering12101131 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Bioengineering, online erschienen am 21.10.2025. Verfasst von Jung-Tae Lee und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Periodontics, One-Stop Specialty Center, Seoul National University Dental Hospital, Seoul 03080, Republic of Korea. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: the Korea Medical Device Development Fund grant funded by the Korea government (the Min-istry of Science and ICT, the Ministry of Trade, Industry and Energy, t…, Korea Medical Device Development Fund, Korea Research Institute of Chemical Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Abutment screw loosening (ASL) is the most frequent mechanical complication in dentistry, leading to prosthetic instability and biological risks. Preload, generated during screw tightening, is critical for maintaining stability but is influenced by torque application, screw geometry, and cyclic loading.

Methods

This in vitro study compared torque loss between two implant systems (Osstem TSIII and KSIII) with different abutment screw designs. Fifty implant–abutment assemblies (n = 5 per torque group) were tested under tightening torques of 20, 25, 30, 35, and 40 Ncm. Initial removal torque (T1) was measured 5 min after tightening, followed by cyclic loading (150 N, 14 Hz, 100,000 cycles). Post-fatigue removal torque (T2) was then recorded, and torque loss rate (%) was calculated. Independent t-tests and a one-way ANOVA were used for statistical analysis.

Results

KSIII consistently exhibited higher T1 and T2 values than TSIII across all torque levels (p < 0.05). The torque loss rate for TSIII ranged from 36.5% (35 Ncm) to 51.8% (20 Ncm), showing a torque-dependent trend (p < 0.05). In contrast, KSIII maintained torque loss rates below 25% at all levels, with no significant differences between torque groups (p > 0.05). On average, torque loss in TSIII was approximately 2.5–3.0 times higher than in KSIII.

Conclusions

The KSIII system demonstrated superior biomechanical stability, with significantly lower torque loss compared with TSIII, independent of torque level. Clinically, these findings suggest that the KSIII system may reduce the incidence of screw loosening and associated complications. A tightening torque of approximately 35 Ncm appeared to provide the most stable preload. Long-term in vivo studies are warranted to confirm these results under clinical conditions.

Schlagwörter

Stichwörter der Autor:innen

abutment screw looseningbiomechanicsdental implantsfatigue loadingimplant–abutment connectionpreloadtorque loss

Autor:innen und Institutionen

  • Department of Periodontics, One-Stop Specialty Center, Seoul National University Dental Hospital, Seoul 03080, Republic of Korea.
  • Research Center for Bio-Based Chemistry, Korea Research Institute of Chemical Technology (KRICT), Ulsan 44429, Republic of Korea.
  • Department of Cogno-Mechatronics Engineering, Pusan National University, Busan 46241, Republic of Korea.
  • Dental Life Science Research Institute, Seoul National University Dental Hospital, Seoul 03080, Republic of Korea.

ORCID-Profile: Jung-Tae Lee, Dong-Wook Han, Bongju Kim

Förderung

  • the Korea Medical Device Development Fund grant funded by the Korea government (the Min-istry of Science and ICT, the Ministry of Trade, Industry and Energy, t…
  • Korea Medical Device Development Fund
  • Korea Research Institute of Chemical Technology

Zitieren

APA

Lee, J. T., Lee, J. C., Han, D. W., & Kim, B. (2025). Effect of Abutment Screw Design on Torque Loss Under Cyclic Fatigue Loading: A Comparison of TSIII and KSIII Implant Systems. Bioengineering, 12(10), 1131. https://doi.org/10.3390/bioengineering12101131

Vancouver

Lee JT, Lee JC, Han DW, Kim B. Effect of Abutment Screw Design on Torque Loss Under Cyclic Fatigue Loading: A Comparison of TSIII and KSIII Implant Systems. Bioengineering. 2025;12(10):1131. doi:10.3390/bioengineering12101131

BibTeX
@article{Lee2025,
  author = {Lee, Jung-Tae and Lee, Jae-Chang and Han, Dong-Wook and Kim, Bongju},
  title = {Effect of Abutment Screw Design on Torque Loss Under Cyclic Fatigue Loading: A Comparison of TSIII and KSIII Implant Systems},
  journal = {Bioengineering},
  year = {2025},
  volume = {12},
  number = {10},
  pages = {1131},
  doi = {10.3390/bioengineering12101131}
}
Kostenlos starten →