EUVIMEDEUVIMED entdecken
FallberichtOpen Access · CC BY-NC-ND 4.0

Removal of a fractured abutment screw within the implant body: A case report

Feng-Chou Cheng, Chen-Ying Wang, Chun-Pin Chiang

Journal of Dental Sciences · 2023-10 · 18 (4) · 1945-1947

DOI: 10.1016/j.jds.2023.06.008 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Journal of Dental Sciences, 2023-10. Verfasst von Feng-Chou Cheng und 2 weiteren Autor:innen. Beteiligte Einrichtung: Chia-Te Dental Clinic, New Taipei City, Taiwan. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 5 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

tps://doi.org/10.1016/j.jds.2023.06. 91-7902/a 2023 Association for Denta e CC BY-NC-ND license (http://creati In addition to the osseointegration failure involving a breakdown of the established osseointegration by periimplantitis, the most common implant failures are mechanical complications such as crown fracture, framework fracture, screw loosening, screw fracture, and fixture fracture due to improper occlusal force. Removing a fractured abutment screw within the implant body is timeconsuming and challenging due to poor visibility of the fractured screw fragment embedded in the implant. Various implant manufacturers have designed screw retrieval instruments to remove fractured screw fragments. A relatively non-invasive technique to remove the fractured screw fragments shall be prioritized for preserving the integrity of the implant without interfering with the implant’s osseointegration. This case report demonstrated the removal of a fractured abutment screw within the implant body by a counter-torque screw-retrieval device and then replacing it with a new abutment screw for prosthesis fabrication later. This 52-year-old male patient consulted the dentist who performed the dental implant for him with the chief complaint of the detached crown of the right maxillary second molar from the dental implant (Fig. 1A). This prosthesis had been used for about seven years. He still wanted to repair the failed prosthesis from the original dentist. Intraoral and radiographic examinations revealed a rootform cylindrical implant (Ankylos , Dentsply Sirona Inc, Charlotte, NC, USA) replacing the right maxillary second molar (Fig. 1B). This dental implant had a fractured

Schlagwörter

Stichwörter der Autor:innen

Counter-torque deviceFractured abutment screwScrew retrieval instruments

Autor:innen und Institutionen

  • Chia-Te Dental Clinic, New Taipei City, Taiwan.
  • School of Life Science, National Taiwan Normal University, Taipei, Taiwan.
  • Science Education Center, National Taiwan Normal University, Taipei, Taiwan.
  • Department of Dentistry, National Taiwan University Hospital, College of Medicine, National Taiwan University, Taipei, Taiwan.
  • School of Dentistry, National Taiwan University, Taipei, Taiwan.
  • Graduate Institute of Clinical Dentistry, School of Dentistry, National Taiwan University, Taipei, Taiwan.

Zitieren

APA

Cheng, F. C., Wang, C. Y., & Chiang, C. P. (2023). Removal of a fractured abutment screw within the implant body: A case report. Journal of Dental Sciences, 18(4), 1945-1947. https://doi.org/10.1016/j.jds.2023.06.008

Vancouver

Cheng FC, Wang CY, Chiang CP. Removal of a fractured abutment screw within the implant body: A case report. Journal of Dental Sciences. 2023;18(4):1945-1947. doi:10.1016/j.jds.2023.06.008

BibTeX
@article{Cheng2023,
  author = {Cheng, Feng-Chou and Wang, Chen-Ying and Chiang, Chun-Pin},
  title = {Removal of a fractured abutment screw within the implant body: A case report},
  journal = {Journal of Dental Sciences},
  year = {2023},
  volume = {18},
  number = {4},
  pages = {1945--1947},
  doi = {10.1016/j.jds.2023.06.008}
}
Kostenlos starten →