Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT
Journal of Nuclear Medicine · 2025-10 · 66 (10) · 1659-1665
DOI: 10.2967/jnumed.125.270078 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine, online erschienen am 28.08.2025. Verfasst von Mostafa Roya und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Motion, Beds. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 21 Literaturangaben.
Abstract (Originalsprache)
Continuous bed motion (CBM) for long-axial-field-of-view PET/CT enables swift total-body examination of patients. However, the sensitivity profile along the axial field of view (AFOV) varies significantly, and its effect on image quality when combined with CBM remains unexplored. This study assesses the effects of ultrahigh sensitivity (UHS) and CBM on recovery coefficients (RCs) and spatial resolution (SR).
Methods
Phantom measurements, performed in static bed and CBM acquisition with different bed speeds (3.5, 7, 14, and 30 mm/s), were reconstructed for both high-sensitivity (HS) and UHS modes, and their quality was assessed using the RC and full width at half maximum of the point-spread function. In addition, 7 clinically referred oncologic patients underwent a combined CBM and UHS [18F]FDG PET/CT scan. Metrics derived from lesions and healthy tissues were compared across acquisition modes.
Results
Mean RC was lower in both UHS ( 0.94) and lesions (r 2 > 0.99) showed excellent correlation. The coefficient of variation was significantly different for the liver.
Conclusion
Elaborate protocols using UHS and CBM could be a step toward swift total-body PET care with only a marginal increase in noise. CBM allows a larger part of the body to be examined and helps to mitigate substantial sensitivity differences along the AFOV, albeit at the cost of slightly lower RC.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
- Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands; c.tsoumpas@umcg.nl.
- Department of Statistics, School of Mathematics, University of Leeds, Leeds, United Kingdom.
- Biomedical Photonic Imaging Group, Faculty of Science and Technology, University of Twente, Enschede, The Netherlands; and.
- Department of Radiology and Nuclear Medicine, Amsterdam UMC, location VUMC, Amsterdam, The Netherlands.
Zitieren
APA
Roya, M., Tsoumpas, C., van Snick, J. H., Dao, V., Willemsen, A. T. M., Slart, R. H. J. A., Boellaard, R., Glaudemans, A. W. J. M., & van Sluis, J. (2025). Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT. Journal of Nuclear Medicine, 66(10), 1659-1665. https://doi.org/10.2967/jnumed.125.270078
Vancouver
Roya M, Tsoumpas C, van Snick JH, Dao V, Willemsen ATM, Slart RHJA, et al. Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT. Journal of Nuclear Medicine. 2025;66(10):1659-1665. doi:10.2967/jnumed.125.270078
BibTeX
@article{Roya2025,
author = {Roya, Mostafa and Tsoumpas, Charalampos and van Snick, Johannes H and Dao, Viet and Willemsen, Antoon T M and Slart, Riemer H J A and Boellaard, Ronald and Glaudemans, Andor W J M and van Sluis, Joyce},
title = {Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra PET/CT},
journal = {Journal of Nuclear Medicine},
year = {2025},
volume = {66},
number = {10},
pages = {1659--1665},
doi = {10.2967/jnumed.125.270078}
}
EUVIMED