Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences

Kazunori Ohashi, Shigehiro Soh, Yoshiomi Sato, Ryoma Miura, Ayumu Funaki, Kaoru Watanabe, Toshiyuki Iwahata, Kazutaka Saito, Kazunori Kubota

Magnetic Resonance in Medical Sciences · 2026 · 25 (4) · n/a

DOI: 10.2463/mrms.tn.2026-0010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Magnetic Resonance in Medical Sciences, 2026. Verfasst von Kazunori Ohashi und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama. Hauptthemen laut MEDLINE: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, Prostate, Prostatic Neoplasms. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aimed to demonstrate the technical feasibility of a dedicated prostate phantom for standardized qualitative evaluation of diffusion-weighted MRI sequences, focusing on signal-to-noise characteristics and image artifacts. A custom-designed phantom simulating prostate tissue, tumor lesions, and rectal gas was used to evaluate 6 diffusion-weighted imaging techniques at 1.5 and 3.0 tesla. Quantitative assessments included signal-to-noise measurements and susceptibility artifacts, while spatial distortion was evaluated by comparison with T2-weighted imaging. Qualitative image quality was assessed using corresponding clinical images as references. Clinical images were evaluated by a radiologist and a urologist, and phantom images were evaluated by 2 radiological technologists. The phantom demonstrated high inter-reader agreement for qualitative image assessment. Clinical image evaluation was performed to confirm whether artifact trends observed in the phantom were similarly observed in clinical images and showed comparable tendencies across sequences. These findings indicate that the proposed prostate phantom provides a practical and reproducible platform for qualitative evaluation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingDiffusion Magnetic Resonance ImagingProstateProstatic NeoplasmsMaleHumansArtifactsSignal-To-Noise RatioReproducibility of ResultsFeasibility StudiesEquipment Design

Stichwörter der Autor:innen

diffusion-weighted imagingimage qualitymagnetic resonance imagingphantomsprostate

Autor:innen und Institutionen

  • Department of Radiological Technology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama.
  • Department of Urology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama.
  • Department of Diagnostic Radiology, Saitama City Hospital, Saitama, Saitama.
  • Department of Radiology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama.

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Ohashi, K., Soh, S., Sato, Y., Miura, R., Funaki, A., Watanabe, K., Iwahata, T., Saito, K., & Kubota, K. (2026). Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences. Magnetic Resonance in Medical Sciences, 25(4), n/a. https://doi.org/10.2463/mrms.tn.2026-0010

Vancouver

Ohashi K, Soh S, Sato Y, Miura R, Funaki A, Watanabe K, et al. Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences. Magnetic Resonance in Medical Sciences. 2026;25(4):n/a. doi:10.2463/mrms.tn.2026-0010

BibTeX
@article{Ohashi2026,
  author = {Ohashi, Kazunori and Soh, Shigehiro and Sato, Yoshiomi and Miura, Ryoma and Funaki, Ayumu and Watanabe, Kaoru and Iwahata, Toshiyuki and Saito, Kazutaka and Kubota, Kazunori},
  title = {Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences},
  journal = {Magnetic Resonance in Medical Sciences},
  year = {2026},
  volume = {25},
  number = {4},
  pages = {n/a},
  doi = {10.2463/mrms.tn.2026-0010}
}
Kostenlos starten →