Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences
Magnetic Resonance in Medical Sciences · 2026 · 25 (4) · n/a
DOI: 10.2463/mrms.tn.2026-0010 ↗Kurz gefasst
Fachartikel in der Zeitschrift Magnetic Resonance in Medical Sciences, 2026. Verfasst von Kazunori Ohashi und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama. Hauptthemen laut MEDLINE: Phantoms, Imaging, Diffusion Magnetic Resonance Imaging, Prostate, Prostatic Neoplasms. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: Japan Society for the Promotion of Science.
Abstract (Originalsprache)
This study aimed to demonstrate the technical feasibility of a dedicated prostate phantom for standardized qualitative evaluation of diffusion-weighted MRI sequences, focusing on signal-to-noise characteristics and image artifacts. A custom-designed phantom simulating prostate tissue, tumor lesions, and rectal gas was used to evaluate 6 diffusion-weighted imaging techniques at 1.5 and 3.0 tesla. Quantitative assessments included signal-to-noise measurements and susceptibility artifacts, while spatial distortion was evaluated by comparison with T2-weighted imaging. Qualitative image quality was assessed using corresponding clinical images as references. Clinical images were evaluated by a radiologist and a urologist, and phantom images were evaluated by 2 radiological technologists. The phantom demonstrated high inter-reader agreement for qualitative image assessment. Clinical image evaluation was performed to confirm whether artifact trends observed in the phantom were similarly observed in clinical images and showed comparable tendencies across sequences. These findings indicate that the proposed prostate phantom provides a practical and reproducible platform for qualitative evaluation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Technology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama.
- Department of Urology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama.
- Department of Diagnostic Radiology, Saitama City Hospital, Saitama, Saitama.
- Department of Radiology, Dokkyo Medical University Saitama Medical Center, Koshigaya Saitama.
Förderung
- Japan Society for the Promotion of Science
Zitieren
APA
Ohashi, K., Soh, S., Sato, Y., Miura, R., Funaki, A., Watanabe, K., Iwahata, T., Saito, K., & Kubota, K. (2026). Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences. Magnetic Resonance in Medical Sciences, 25(4), n/a. https://doi.org/10.2463/mrms.tn.2026-0010
Vancouver
Ohashi K, Soh S, Sato Y, Miura R, Funaki A, Watanabe K, et al. Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences. Magnetic Resonance in Medical Sciences. 2026;25(4):n/a. doi:10.2463/mrms.tn.2026-0010
BibTeX
@article{Ohashi2026,
author = {Ohashi, Kazunori and Soh, Shigehiro and Sato, Yoshiomi and Miura, Ryoma and Funaki, Ayumu and Watanabe, Kaoru and Iwahata, Toshiyuki and Saito, Kazutaka and Kubota, Kazunori},
title = {Development of a Dedicated Prostate Phantom for Technical Evaluation of Diffusion-weighted MR Imaging Sequences},
journal = {Magnetic Resonance in Medical Sciences},
year = {2026},
volume = {25},
number = {4},
pages = {n/a},
doi = {10.2463/mrms.tn.2026-0010}
}
EUVIMED