The Edwards Acumen IQ system using peripheral arterial catheter-based waveforms to estimate cardiac output is not accurate as compared to thermodilution in dogs

Kristin M Zersen iD, Gregg M Griffenhagen iD, Eric L Monnet iD

American Journal of Veterinary Research · 2024

DOI: 10.2460/ajvr.23.11.0249 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Veterinary Research, 2024. Verfasst von Kristin M Zersen und 2 weiteren Autor:innen. Hauptthemen laut MEDLINE: Thermodilution, Dog Diseases. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE To assess the agreement between cardiac output (CO) estimated via evaluation of the arterial pressure waveform by a novel monitoring system (Edwards Acumen IQ sensor and HemoSphere Advanced Monitor Platform [HS-IQ]; Edwards LifeSciences) and measured by thermodilution (TD) in anesthetized, normovolemic, and hypovolemic dogs. To assess the agreement between the HS-IQ CO measurements in the radial artery and dorsal metatarsal artery. ANIMALS 8 purpose-bred Beagles. METHODS Dogs were placed under general anesthesia. CO was measured via TD and via the HS-IQ at radial and dorsal metatarsal arterial catheters. CO measurements were obtained at 4 time points including normovolemic and multiple hypovolemic states. Paired measurements of CO were evaluated via the method of Bland and Altman with acceptable limits of agreement (LOA) defined as < 30%. RESULTS A total of 24 (dorsal metatarsal) and 21 (radial) paired measurements were collected in 8 dogs. The overall bias (CI) for comparison of TD to radial arterial HS-IQ CO measurements was −0.09 L/min. LOA and proportional LOA were −2.66 to 2.49 L/min and −140.72% to 104.94%. The overall bias (CI) for comparison of TD to dorsal metatarsal arterial HS-IQ CO measurements was −0.26 L/min. LOA and proportional LOA were −2.76 to 2.24 L/min and −135.96% to 93.25%. The overall proportional error for radial arterial was −17.9% and for dorsal metatarsal was −21.4%. CLINICAL RELEVANCE CO measurements with the HS-IQ were easy to obtain but did not produce results within a clinically acceptable range for either measurement site, with a very wide LOA. The CO estimations from the HS-IQ are not appropriate for clinical use at this time.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ThermodilutionDog DiseasesDogsAnimalsHypovolemiaMonitoring, IntraoperativeCardiac OutputArteriesCatheters, IndwellingReproducibility of Results

Stichwörter der Autor:innen

arterial cathetercardiac outputhypovolemianormovolemiathermodilution

Autor:innen und Institutionen

ORCID-Profile: Kristin M Zersen, Gregg M Griffenhagen, Eric L Monnet

Zitieren

APA

Zersen, K. M., Griffenhagen, G. M., & Monnet, E. L. (2024). The Edwards Acumen IQ system using peripheral arterial catheter-based waveforms to estimate cardiac output is not accurate as compared to thermodilution in dogs. American Journal of Veterinary Research. https://doi.org/10.2460/ajvr.23.11.0249

Vancouver

Zersen KM, Griffenhagen GM, Monnet EL. The Edwards Acumen IQ system using peripheral arterial catheter-based waveforms to estimate cardiac output is not accurate as compared to thermodilution in dogs. American Journal of Veterinary Research. 2024. doi:10.2460/ajvr.23.11.0249

BibTeX
@article{Zersen2024,
  author = {Zersen, Kristin M and Griffenhagen, Gregg M and Monnet, Eric L},
  title = {The Edwards Acumen IQ system using peripheral arterial catheter-based waveforms to estimate cardiac output is not accurate as compared to thermodilution in dogs},
  journal = {American Journal of Veterinary Research},
  year = {2024},
  doi = {10.2460/ajvr.23.11.0249}
}
Kostenlos starten →