Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery

Yusuke Takei, Michio Kumagai, Manami Suzuki, Sakura Mori, Yuna Sato, Toru Tamii, Akane Tamii, Ako Saito, Yuko Ogata, Yu Kaiho, Hiroaki Toyama, Yutaka Ejima und 1 weitere

Journal of Cardiothoracic and Vascular Anesthesia · 2023-07 · 37 (7) · 1143-1151

DOI: 10.1053/j.jvca.2023.03.019 ↗

Kurz gefasst

Beobachtungsstudie in der Zeitschrift Journal of Cardiothoracic and Vascular Anesthesia, 2023-07. Verfasst von Yusuke Takei und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology and Perioperative Medicine, Tohoku University Graduate School of Medicine, Sendai, Japan. Electronic address: ytakei@med…. Hauptthemen laut MEDLINE: Monitoring, Intraoperative, Cardiac Surgical Procedures. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 38 Literaturangaben. Gefördert von: Japan Society for the Promotion of Science.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: The clinical use of less-invasive devices that calculate the cardiac output from arterial pressure waveform is increasing. The authors aimed to evaluate the accuracy and characteristics of the systemic vascular resistance index (SVRI) of the cardiac index measured by 2 less-invasive devices, fourth-generation FloTrac (CIFT) and LiDCOrapid (CILR), compared with the intermittent thermodilution technique, using a pulmonary artery catheter (CITD). DESIGN: This was a prospective observational study. SETTING: This study was conducted at a single university hospital. PARTICIPANTS: Twenty-nine adult patients undergoing elective cardiac surgery. INTERVENTIONS: Elective cardiac surgery was used as an intervention. MEASUREMENTS AND MAIN RESULTS: Hemodynamic parameters, CIFT, CILR, and CITD, were measured after the induction of general anesthesia, at the start of cardiopulmonary bypass, after completion of weaning from cardiopulmonary bypass, 30 minutes after weaning, and at sternal closure (135 measurements in total). The CIFT and CILR had moderate correlations with CITD (r = 0.62 and 0.58, respectively). Compared with CITD, CIFT, and CILR had a bias of -0.73 and -0.61 L/min/m2, limit of agreement of -2.14-to-0.68 L/min/m2 and -2.42-to-1.20 L/min/m2, and percentage error of 39.9% and 51.2%, respectively. Subgroup analysis for evaluating SVRI characteristics showed that the percentage errors of CIFT and CILR were 33.9% and 54.5% in low SVRI ( 1,800 dyne·s/cm5/m2), respectively. CONCLUSIONS: The accuracy of CIFT or CILR was not clinically acceptable for cardiac surgery. Fourth-generation FloTrac was unreliable in high SVRI. LiDCOrapid was inaccurate across a broad range of SVRI, and minimally affected by SVRI.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monitoring, IntraoperativeCardiac Surgical ProceduresAdultHumansCardiac OutputVascular ResistanceHemodynamicsThermodilutionReproducibility of Results

Stichwörter der Autor:innen

FloTracLiDCOrapidarterial pressure-based cardiac outputcardiac surgerysystemic vascular resistance index

Autor:innen und Institutionen

  • Department of Anesthesiology and Perioperative Medicine, Tohoku University Graduate School of Medicine, Sendai, Japan. Electronic address: ytakei@med.tohoku.ac.jp.
  • Department of Anesthesiology and Perioperative Medicine, Tohoku University Graduate School of Medicine, Sendai, Japan.
  • Department of Anesthesiology, Osaki Citizen Hospital, Osaki, Japan.
  • Department of Anesthesiology, Sendai Medical Center, Sendai, Japan.

Förderung

  • Japan Society for the Promotion of Science

Zitieren

APA

Takei, Y., Kumagai, M., Suzuki, M., Mori, S., Sato, Y., Tamii, T., Tamii, A., Saito, A., Ogata, Y., Kaiho, Y., Toyama, H., Ejima, Y., & Yamauchi, M. (2023). Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery. Journal of Cardiothoracic and Vascular Anesthesia, 37(7), 1143-1151. https://doi.org/10.1053/j.jvca.2023.03.019

Vancouver

Takei Y, Kumagai M, Suzuki M, Mori S, Sato Y, Tamii T, et al. Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery. Journal of Cardiothoracic and Vascular Anesthesia. 2023;37(7):1143-1151. doi:10.1053/j.jvca.2023.03.019

BibTeX
@article{Takei2023,
  author = {Takei, Yusuke and Kumagai, Michio and Suzuki, Manami and Mori, Sakura and Sato, Yuna and Tamii, Toru and Tamii, Akane and Saito, Ako and Ogata, Yuko and Kaiho, Yu and Toyama, Hiroaki and Ejima, Yutaka and Yamauchi, Masanori},
  title = {Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery},
  journal = {Journal of Cardiothoracic and Vascular Anesthesia},
  year = {2023},
  volume = {37},
  number = {7},
  pages = {1143--1151},
  doi = {10.1053/j.jvca.2023.03.019}
}
Kostenlos starten →