EUVIMEDOpdag EUVIMED
MetaanalyseSystematische ÜbersichtOpen access

Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis

Sanidhya Singh iD, Miles Romney Bennett iD, Chen Chen iD, Sooyoon Shin iD, Hamid Ghanbari iD, Benjamin W Nelson iD

Journal of Medical Internet Research · 2024-10-10 · 26 · e62769

DOI: 10.2196/62769 ↗

At a glance

Metaanalyse in Journal of Medical Internet Research, published online on 2024-10-10. Written by Sanidhya Singh and 5 co-authors. Institution: University of Michigan Medical School, Ann Arbor, MI, United States. Major MEDLINE topics: Oximetry, Photoplethysmography, Skin Pigmentation, Wearable Electronic Devices. Cited 38 times so far (Crossref); 75 references. The full text is freely available in PubMed Central.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO2) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities.

Objective

This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO2 and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO2 and PR according to skin pigmentation.

Methods

We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO2-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO2 accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or –5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI.

Results

For SpO2, accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI –0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from –16.02 to 13.54, from –18.62 to 16.84, and from –33.69 to 32.54, and pooled mean biases were –1.24 (95% CI –5.31 to 2.83) bpm, –0.89 (95% CI –3.70 to 1.93) bpm, and –0.57 (95% CI –9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively.

Conclusions

SpO2 and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO2 for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.

Subjects

MeSH terms (MEDLINE)

OximetryPhotoplethysmographySkin PigmentationWearable Electronic DevicesHumansHeart RateOxygen

Author keywords

arterial blood gasbiasdigital technologyphotoplethysmographypulse oximetryskin pigmentationskin tone

Substances

Oxygen

Authors and institutions

  • University of Michigan Medical School, Ann Arbor, MI, United States.
  • Verily Life Sciences LLC, South San Francisco, CA, United States.
  • Division of Digital Psychiatry, Department of Psychiatry, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA, United States.

ORCID profiles: Sanidhya Singh, Miles Romney Bennett, Chen Chen, Sooyoon Shin, Hamid Ghanbari, Benjamin W Nelson

Cite

APA

Singh, S., Bennett, M. R., Chen, C., Shin, S., Ghanbari, H., & Nelson, B. W. (2024). Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis. Journal of Medical Internet Research, 26, e62769. https://doi.org/10.2196/62769

Vancouver

Singh S, Bennett MR, Chen C, Shin S, Ghanbari H, Nelson BW. Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis. Journal of Medical Internet Research. 2024;26:e62769. doi:10.2196/62769

BibTeX
@article{Singh2024,
  author = {Singh, Sanidhya and Bennett, Miles Romney and Chen, Chen and Shin, Sooyoon and Ghanbari, Hamid and Nelson, Benjamin W},
  title = {Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis},
  journal = {Journal of Medical Internet Research},
  year = {2024},
  volume = {26},
  pages = {e62769},
  doi = {10.2196/62769}
}
Start for free →