Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study
Sensors · 2026
DOI: 10.3390/s26041201 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Sensors, 2026. Verfasst von Patrick van den Berge und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, University Medical Center Utrecht, Utrecht University, 3508 TC Utrecht, The Netherlands. Hauptthemen laut MEDLINE: Photoplethysmography, Vital Signs, Wearable Electronic Devices, Wireless Technology. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: FastFocus BV., FastFocus B.V..
Abstract (Originalsprache)
(1)
Background
Evidence on the clinical accuracy of wireless photoplethysmography (PPG)-based vital sign monitoring is limited. This study evaluated the accuracy, technical performance, and patient comfort of a novel PPG-based earsensor for measuring oxygen saturation (SpO2), pulse rate (PR), and respiratory rate (RR) in postoperative patients. (2)
Methods
In this observational method comparison study, SpO2, PR, and RR were simultaneously recorded using the earsensor and compared with continuous monitoring in patients admitted overnight to the post-anesthesia care unit. Outcome measures were bias, 95% limits of agreement (LoA), and average root mean square (ARMS). Technical performance was evaluated by data loss and data gap duration. Patient comfort was assessed using a questionnaire. (3)
Results
Twenty-one patients contributed to 264 h of data. Bias was 1.7% for SpO2 (ARMS 2.4%; LoA −1.8% to 5.1%), 1.2 bpm for PR (ARMS 3.9 bpm; LoA –6.1 to 8.4 bpm), and 0.3 brpm for RR (ARMS 4.4 brpm; LoA –8.4 to 8.9 brpm). Overall, data loss was 42% for SpO2, 33% for RR, and 29% for PR; most data gaps were under 30 min. Patient-reported comfort was high (77%). (4)
Conclusions
The earsensor accurately measured SpO2 and PR. RR accuracy was outside the predefined criteria. Despite substantial data loss, patient comfort was high, supporting the potential of PPG-based sensors for unobtrusive vital sign trend monitoring in low-acuity settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Anesthesiology, University Medical Center Utrecht, Utrecht University, 3508 TC Utrecht, The Netherlands.
- FastFocus B.V., 3481 LT Harmelen, The Netherlands.
- Department of Technical Medicine, University of Twente, 7500 AE Enschede, The Netherlands.
- Department of Internal Medicine, University Medical Center Utrecht, Utrecht University, 3508 TC Utrecht, The Netherlands.
ORCID-Profile: Lianne Zevenbergen, Martine J M Breteler
Förderung
- FastFocus BV.
- FastFocus B.V.
Zitieren
APA
van den Berge, P., van Loon, K., Zevenbergen, L., van den Heuvel, P. A., & Breteler, M. J. M. (2026). Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study. Sensors. https://doi.org/10.3390/s26041201
Vancouver
van den Berge P, van Loon K, Zevenbergen L, van den Heuvel PA, Breteler MJM. Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study. Sensors. 2026. doi:10.3390/s26041201
BibTeX
@article{vandenBerge2026,
author = {van den Berge, Patrick and van Loon, Kim and Zevenbergen, Lianne and van den Heuvel, Pascalle A and Breteler, Martine J M},
title = {Continuous Vital Signs Monitoring with a Wireless and Wearable Earsensor in Surgical Patients: A Clinical Validation Study},
journal = {Sensors},
year = {2026},
doi = {10.3390/s26041201}
}
EUVIMED