Research on Structural Optimization of Interventional Micro-Axial Blood Pump Based on Nondominated Sorting Genetic Algorithm-II
Journal of Biomechanical Engineering · 01.08.2026 · 148 (8)
DOI: 10.1115/1.4071621 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Biomechanical Engineering, online erschienen am 03.06.2026. Verfasst von Gang Cheng und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Energy and Power Engineering, University of Shanghai for Science and Technology, Shanghai 200039, China. Hauptthemen laut MEDLINE: Genetic Algorithms, Heart-Assist Devices. Die Arbeit nennt 33 Literaturangaben.
Abstract (Originalsprache)
This study presents an optimization approach for interventional micro-axial blood pumps, aimed at addressing the rising burden of cardiovascular diseases and the limited availability of heart transplant donors in China. The methodology integrates design of experiments, computational fluid dynamics (CFD) analysis, and response surface methodology (RSM), with the Non-dominated Sorting Genetic Algorithm-II (NSGA-II) multi-objective genetic algorithm employed to optimize impeller geometry. The primary objectives were to enhance hydraulic efficiency and improve hemocompatibility. The optimized blood pump model demonstrated notable performance improvements across various flow conditions: at a low flow rate of 1 L/min, the pump head increased by 2.97% and the hemolysis index decreased by 12.04%; at the design point, head improved by 5.22% while hemolysis was reduced by 11.71%; under high-flow conditions (4 L/min), head increased by 8.5% and hemolysis decreased by 12.57%. These enhancements contribute to higher energy efficiency and lower hemolytic risk, thereby improving the safety and reliability of the device in clinical applications. The findings provide a robust foundation for future advancements in blood pump design and optimization.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- School of Energy and Power Engineering, University of Shanghai for Science and Technology, Shanghai 200039, China.
- University of Shanghai for Science and Technology.
- Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai Institute of Cardiovascular Diseases, Shanghai 200032, China.
ORCID-Profile: Gang Cheng, Ying Wang
Zitieren
APA
Cheng, G., Geng, X., Ma, J., & Wang, Y. (2026). Research on Structural Optimization of Interventional Micro-Axial Blood Pump Based on Nondominated Sorting Genetic Algorithm-II. Journal of Biomechanical Engineering, 148(8). https://doi.org/10.1115/1.4071621
Vancouver
Cheng G, Geng X, Ma J, Wang Y. Research on Structural Optimization of Interventional Micro-Axial Blood Pump Based on Nondominated Sorting Genetic Algorithm-II. Journal of Biomechanical Engineering. 2026;148(8). doi:10.1115/1.4071621
BibTeX
@article{Cheng2026,
author = {Cheng, Gang and Geng, Xuesong and Ma, Jianying and Wang, Ying},
title = {Research on Structural Optimization of Interventional Micro-Axial Blood Pump Based on Nondominated Sorting Genetic Algorithm-II},
journal = {Journal of Biomechanical Engineering},
year = {2026},
volume = {148},
number = {8},
doi = {10.1115/1.4071621}
}