Impact of extracorporeal blood pump gap sizes on the performance and hemocompatibility under off‐design operation

Lars Fischer iD, Sebastian V Jansen, Ulrich Steinseifer, Ifan Yen, Po-Lin Hsu, Michael Neidlin

Artificial Organs · 2025

DOI: 10.1111/aor.14875 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Artificial Organs, 2025. Verfasst von Lars Fischer und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany. Hauptthemen laut MEDLINE: Heart-Assist Devices. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: magAssist Co., Ltd.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Hemocompatibility remains the dominant challenge in rotary blood pumps, and more information on the relationship between individual pump design features, hemodynamics, and blood trauma in various operation conditions is necessary. The study evaluated the variation of gap sizes in extracorporeal blood pumps concerning their influence on blood compatibility, particularly during off‐design conditions.

Methods

We developed a parametric generic blood pump framework for in‐silico and in‐vitro design feature analysis. Thirty‐six designs with varying axial and radial gap sizes between 0.5 mm and 3 mm were generated. CFD was applied to calculate and compare device hemodynamics and evaluate the performance and hemocompatibility during off‐design and target operation conditions. The following quantities were analyzed: pressure difference, hemolysis potential, residence times, hydraulic efficiency, and recirculation ratio.

Results

The in‐vitro prototype showed excellent agreement with in‐silico predictions regarding hydraulic performance ( R 2 = 0.996 with a RMSE = 2.07). Our results show a modest impact of gap size variations ±10% on key metrics. Domain‐resolved analyses revealed a significant contribution of the gap regions to the device's overall hemolytic performance, with an increasing contribution for off‐design flow rates. Overall elevated hemolysis levels were identified if at least one gap size was held minimal.

Conclusions

We introduced and showed the feasibility of a parametric rotary blood pump framework to systematically investigate design feature impact. Results suggest, larger and uniformly sized gaps being overall beneficial regarding hemocompatibility.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart-Assist DevicesHemolysisHumansHemodynamicsMaterials TestingEquipment DesignComputer SimulationModels, CardiovascularProsthesis DesignHydrodynamics

Stichwörter der Autor:innen

CFDblood damageblood pumpgapsoff design

Autor:innen und Institutionen

  • Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.
  • magAssist Co., Ltd., Suzhou, China.

ORCID-Profile: Lars Fischer

Förderung

  • magAssist Co., Ltd

Zitieren

APA

Fischer, L., Jansen, S. V., Steinseifer, U., Yen, I., Hsu, P. L., & Neidlin, M. (2025). Impact of extracorporeal blood pump gap sizes on the performance and hemocompatibility under off‐design operation. Artificial Organs. https://doi.org/10.1111/aor.14875

Vancouver

Fischer L, Jansen SV, Steinseifer U, Yen I, Hsu PL, Neidlin M. Impact of extracorporeal blood pump gap sizes on the performance and hemocompatibility under off‐design operation. Artificial Organs. 2025. doi:10.1111/aor.14875

BibTeX
@article{Fischer2025,
  author = {Fischer, Lars and Jansen, Sebastian V and Steinseifer, Ulrich and Yen, Ifan and Hsu, Po-Lin and Neidlin, Michael},
  title = {Impact of extracorporeal blood pump gap sizes on the performance and hemocompatibility under off‐design operation},
  journal = {Artificial Organs},
  year = {2025},
  doi = {10.1111/aor.14875}
}
Kostenlos starten →