Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates
Artificial Organs · 2025
DOI: 10.1111/aor.14899 ↗Kurz gefasst
Fachartikel in der Zeitschrift Artificial Organs, 2025. Verfasst von Mathieu Specklin und 5 weiteren Autor:innen. Beteiligte Einrichtung: Cnam, Arts et Metiers Institute of Technology, LIFSE, Paris, France. Hauptthemen laut MEDLINE: Heart-Assist Devices, Heart Failure. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
An innovative solution has been recently proposed for the treatment of heart failure with preserved ejection fraction (HFpEF), using a centrifugal mechanical circulatory support (MCS) device. We sought firstly to assess the hemocompatibility of the proposed device. HFpEF treatment requires the blood pump to operate at low blood flow rate (0.05–0.5 L/min). Given high blood trauma expected in these conditions, we sought secondly to investigate design improvements in order to reduce this risk.
Method
Computational fluid dynamics was used to analyze the blood fluid filled within the centrifugal pump and to estimate the blood trauma due to its operation. This assessment is based on the computation of integrated quantities such as the hemolysis index and the turbulent dissipation energy.
Results
The hemolysis index associated with the present device is comparable to the figures from HVAD. With the proposed pump, for instance, an index of 0.0036 is obtained at 1 L/min and 3000 rpm. Using thinner blades for the impeller allows a 12% reduction of the hemolysis index in average, while reducing its diameter leads to an index 2.8 folds lower at 0.5 L/min.
Conclusion
The present investigation shows the promising hemocompatibility of our centrifugal pump. Concerns about very high hemolysis generated at low flow rates could be overcome by reducing the impeller diameter. Experimental validations are planned to support our findings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cnam, Arts et Metiers Institute of Technology, LIFSE, Paris, France.
- Arts et Metiers Institute of Technology, Cnam, LIFSE, F-75013, Paris, France.
- Department of Cardiovascular and Thoracic Surgery, Institute of Cardiology, Pitié Salpêtrière Hospital, Assistance Publique Hôpitaux de Paris (AP-HP), Sorbonne University, Paris, France.
ORCID-Profile: Mathieu Specklin, Navideh Abbasnezhad
Zitieren
APA
Specklin, M., Marcel, L., Abbasnezhad, N., Bakir, F., Kouidri, S., & Danial, P. (2025). Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates. Artificial Organs. https://doi.org/10.1111/aor.14899
Vancouver
Specklin M, Marcel L, Abbasnezhad N, Bakir F, Kouidri S, Danial P. Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates. Artificial Organs. 2025. doi:10.1111/aor.14899
BibTeX
@article{Specklin2025,
author = {Specklin, Mathieu and Marcel, Louis and Abbasnezhad, Navideh and Bakir, Farid and Kouidri, Smaine and Danial, Pichoy},
title = {Improving the hemocompatibility of centrifugal mechanical circulatory support devices at low flow rates},
journal = {Artificial Organs},
year = {2025},
doi = {10.1111/aor.14899}
}
EUVIMED