Continuous carbon dioxide monitoring in the exhaled breath of mechanically ventilated rats

Christin Wenzel iD, Silke Borgmann, Bernd Flamm, Lea Kuhn, Stefan Schumann, Johannes Schmidt, Johannes Spaeth, Sashko Spassov iD

Experimental Physiology · 2025

DOI: 10.1113/ep093058 ↗

At a glance

Journal article in Experimental Physiology, 2025. Written by Christin Wenzel and 7 co-authors. Institution: Department of Anesthesiology and Critical Care, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germa…. Major MEDLINE topics: Carbon Dioxide, Respiration, Artificial, Exhalation. 17 references. The full text is freely available (open access, CC BY 4.0). Funded by: DFG grant SCHU.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

In small animals, carbon dioxide monitoring is either limited by the need to take blood samples for gas analysis, or it interferes with respiratory efficiency and lung mechanics analysis. We introduced a novel approach for continuous monitoring of CO 2 in the expiratory limb of the breathing circuit. The relevance of the method is assessed by CO 2 measurements at different respiratory settings. Rats were ventilated with a tidal volume ( V T ) of 8 mL kg −1 . Respiratory rates were adjusted to achieve arterial CO 2 partial pressure () between 35 and 45 mmHg. We measured partial pressure of CO 2 in the expiratory limb of the breathing circuit (exCO 2 ). exCO 2 values were compared to from blood gas analysis. The agreement between the two measurements was assessed by correlation and Bland–Altman analysis. The validity of the novel approach was established through additional experimental runs with V T of 7 or 6 mL kg −1 , where the respiratory rate was set in accordance to exCO 2 . Measurements of exCO 2 reflected with high correlation ( R 2 = 0.8658). Bland–Altman analysis showed high agreement between the two measurements. Respiratory rate setting guided by exCO 2 (39 ± 2 or 40 ± 2 mmHg during ventilation with V T of 7 or 6 mL kg −1 , respectively) was appropriate to maintain (41 ± 2 mmHg for both V T ). Measurement of exCO 2 provides a robust estimate of in anaesthetized small animals during mechanical ventilation. Continuous monitoring of CO 2 in the exhaled breath could be used to guide mechanical ventilation settings.

Subjects

MeSH terms (MEDLINE)

Carbon DioxideRespiration, ArtificialExhalationAnimalsRatsMaleTidal VolumeMonitoring, PhysiologicBlood Gas AnalysisBreath TestsRespiratory MechanicsRats, WistarPartial PressureRats, Sprague-Dawley

Author keywords

CO2 partial pressurecarbon dioxideend‐tidalmechanical ventilationmonitoringrespiratory efficiencysmall animals

Substances

Carbon Dioxide

Authors and institutions

  • Department of Anesthesiology and Critical Care, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Freiburg, Germany.

ORCID profiles: Christin Wenzel, Sashko Spassov

Funding

  • DFG grant SCHU

Cite

APA

Wenzel, C., Borgmann, S., Flamm, B., Kuhn, L., Schumann, S., Schmidt, J., Spaeth, J., & Spassov, S. (2025). Continuous carbon dioxide monitoring in the exhaled breath of mechanically ventilated rats. Experimental Physiology. https://doi.org/10.1113/ep093058

Vancouver

Wenzel C, Borgmann S, Flamm B, Kuhn L, Schumann S, Schmidt J, et al. Continuous carbon dioxide monitoring in the exhaled breath of mechanically ventilated rats. Experimental Physiology. 2025. doi:10.1113/ep093058

BibTeX
@article{Wenzel2025,
  author = {Wenzel, Christin and Borgmann, Silke and Flamm, Bernd and Kuhn, Lea and Schumann, Stefan and Schmidt, Johannes and Spaeth, Johannes and Spassov, Sashko},
  title = {Continuous carbon dioxide monitoring in the exhaled breath of mechanically ventilated rats},
  journal = {Experimental Physiology},
  year = {2025},
  doi = {10.1113/ep093058}
}
Start for free →