Effect of increased resistance on dynamic compliance assessed by two clinical monitors during volume-controlled ventilation: A test-lung study
Veterinary Anaesthesia and Analgesia · 2024
DOI: 10.1016/j.vaa.2024.03.003 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Anaesthesia and Analgesia, 2024. Verfasst von Daniel Ben-Aderet und 2 weiteren Autor:innen. Beteiligte Einrichtung: William Pritchard Veterinary Medical Teaching Hospital, School of Veterinary Medicine, University of California Davis, Davis, CA, USA. Hauptthemen laut MEDLINE: Respiration, Artificial. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: School of Veterinary Medicine, University of California, Davis, University of California Davis Center for Companion Animal Health.
Abstract (Originalsprache)
OBJECTIVE: To evaluate the effect of increased respiratory system resistance (RRS) on dynamic compliance (Cdyn) assessed by the NM3 monitor (Cdyn(NM3)) and the E-CAiOV module (Cdyn(ECAiOV)). STUDY DESIGN: Prospective laboratory study. METHODS: A training test lung (TTL) simulated the mechanical ventilation of a mammal with 50 and 300 mL tidal volumes in three conditions of RRS [normal (RBL), moderately increased (R1) and severely increased (R2)] and a wide range of clinically relevant Cdyn. Simulations at increased RRS were paired with simulations at RBL with the same static compliance for comparisons. Pearson's correlation coefficient and concordance correlation coefficient between the measurements at RBL with the ones with increased RRS were calculated. Bland-Altman plots were also used to evaluate the agreement of Cdyn(ECAiOV) and Cdyn(NM3) at RBL (control values) with their paired values at R1 and R2. Relative bias and limits of agreement (LOAs) were calculated and LOAs larger than 30% were considered unacceptable. Trending ability of Cdyn(NM3) and Cdyn(ECAiOV) were evaluated by polar plots. Values of p < 0.05 were considered significant. RESULTS: The effect of increased RRS was more pronounced for Cdyn(ECAiOV) than for Cdyn(NM3). Unacceptable agreement was only observed in Cdyn(NM3) at R2 in the 300 mL simulation (bias = -18.3% and lower LOA = -45%). For Cdyn(ECAiOV), agreement was unacceptable for all tested RRS in both simulations, being the worst at R2 in the 300 mL simulation (bias = -54.7% and lower LOA = -100.2%). Both levels of increased RRS caused poor trending ability for Cdyn(ECAiOV), whereas the same effect was only observed for Cdyn(NM3) at R2. CONCLUSIONS AND CLINICAL RELEVANCE: In the presence of increased RRS, Cdyn estimated by the NM3 monitor presented better capability to distinguish between changes in RRS from changes in respiratory system compliance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- William Pritchard Veterinary Medical Teaching Hospital, School of Veterinary Medicine, University of California Davis, Davis, CA, USA.
- Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California Davis, Davis, CA, USA. Electronic address: jhsoares@ucdavis.edu.
- Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California Davis, Davis, CA, USA.
Förderung
- School of Veterinary Medicine, University of California, Davis
- University of California Davis Center for Companion Animal Health
Zitieren
APA
Ben-Aderet, D., Soares, J. H., & Bueno, M. L. (2024). Effect of increased resistance on dynamic compliance assessed by two clinical monitors during volume-controlled ventilation: A test-lung study. Veterinary Anaesthesia and Analgesia. https://doi.org/10.1016/j.vaa.2024.03.003
Vancouver
Ben-Aderet D, Soares JH, Bueno ML. Effect of increased resistance on dynamic compliance assessed by two clinical monitors during volume-controlled ventilation: A test-lung study. Veterinary Anaesthesia and Analgesia. 2024. doi:10.1016/j.vaa.2024.03.003
BibTeX
@article{BenAderet2024,
author = {Ben-Aderet, Daniel and Soares, João Hn and Bueno, Melissa Lc},
title = {Effect of increased resistance on dynamic compliance assessed by two clinical monitors during volume-controlled ventilation: A test-lung study},
journal = {Veterinary Anaesthesia and Analgesia},
year = {2024},
doi = {10.1016/j.vaa.2024.03.003}
}
EUVIMED