Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes
Veterinary Ophthalmology · 2027-01 · 30 (1)
DOI: 10.1111/vop.70300 ↗Kurz gefasst
Fachartikel in der Zeitschrift Veterinary Ophthalmology, online erschienen am 13.11.2026. Verfasst von Haobo Zhao und 3 weiteren Autor:innen. Beteiligte Einrichtung: The School of Mechanical Engineering, Zhejiang Sci-Tech University, Hangzhou, China. Hauptthemen laut MEDLINE: Intraocular Pressure, Tonometry, Ocular. Die Arbeit nennt 24 Literaturangaben. Gefördert von: Zhejiang Sci-Tech University.
Abstract (Originalsprache)
Objective
To validate the accuracy of the iFalcon V700 rebound tonometer in measuring intraocular pressure (IOP) in enucleated normal rat eyes. Animals Studied Ten freshly enucleated normal rat eyes were tested within 12 h postmortem. Procedures Enucleated rat eyes were fixed to the front of the device. IOP measurements started at 5 mmHg and were gradually increased to 95 mmHg by mechanical pressure from the vitreous side toward the anterior chamber. The iFalcon V700 rebound tonometer measured central corneal IOP in rodent mode, and results were compared with the iCare TonoLab using linear regression analysis and Bland–Altman plots.
Results
Measurements from the iFalcon V700 correlated strongly with the iCare TonoLab ( r 2 = 0.98). Bland–Altman analysis showed a mean bias of −1.56 mmHg, with 95% limits of agreement from −6.71 to +3.59 mmHg, and slight underestimation in the moderate IOP range. In the 5–10 mmHg range, multiple attempts were needed to get valid readings with the iFalcon V700.
Conclusions
The iFalcon V700 enables accurate central‐corneal IOP measurement (5–95 mmHg) in enucleated normal rat eyes and shows good agreement with iCare TonoLab; repeated measurements at identical pressure levels improve precision. This study has limitations: inherent errors from iCare TonoLab cannot be excluded, and mechanical compression precludes testing at extremely low IOP (0–5 mmHg). Despite its promising potential for rodent tonometry, further systematic in vivo validation is required prior to its widespread, reliable use in rat ophthalmic research.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- The School of Mechanical Engineering, Zhejiang Sci-Tech University, Hangzhou, China.
- Pai Hong Tai Pet Hospital (Huancheng Main Hospital), Hangzhou, China.
ORCID-Profile: Ji Fu
Förderung
- Zhejiang Sci-Tech University
Zitieren
APA
Zhao, H., Su, Y., Yang, L., & Fu, J. (2027). Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes. Veterinary Ophthalmology, 30(1). https://doi.org/10.1111/vop.70300
Vancouver
Zhao H, Su Y, Yang L, Fu J. Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes. Veterinary Ophthalmology. 2027;30(1). doi:10.1111/vop.70300
BibTeX
@article{Zhao2027,
author = {Zhao, Haobo and Su, Yunhao and Yang, Lihui and Fu, Ji},
title = {Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes},
journal = {Veterinary Ophthalmology},
year = {2027},
volume = {30},
number = {1},
doi = {10.1111/vop.70300}
}
EUVIMED