EUVIMEDEUVIMED entdecken
VergleichsstudieValidierungsstudieOpen Access

Ex Vivo Validation of the TONOVET Plus and TONOVET Pro in Cynomolgus Monkeys ( Macaca fascicularis ): Comparison of Canine and Lapine Calibration Modes

Wontae Kim iD, Seonmi Kang iD, Jiwon Lee iD, Jun Young Lee iD, Kangmoon Seo iD

Veterinary Ophthalmology · 2026-09 · 29 (5)

DOI: 10.1111/vop.70265 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Veterinary Ophthalmology, online erschienen am 07.09.2026. Verfasst von Wontae Kim und 4 weiteren Autor:innen. Beteiligte Einrichtung: KeyPrime Research, Cheongju-si, Chungcheongbuk-do, Korea. Hauptthemen laut MEDLINE: Macaca fascicularis, Intraocular Pressure, Tonometry, Ocular. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To evaluate the accuracy of the TONOVET Plus (TVP) and TONOVET Pro (TVPR) in measuring intraocular pressure (IOP) in ex vivo cynomolgus monkey eyes compared to manometry, and to determine the accuracy of Canine and Lapine calibration modes: Animals Studied Five enucleated eyes from three cynomolgus monkeys. Procedures The anterior chambers were cannulated to regulate IOP manometrically from 5 to 80 mmHg. Agreement between tonometric and manometric readings was analyzed using a linear mixed‐effects model (LMM), Lin's concordance correlation coefficient (CCC), and repeated‐measures Bland–Altman analysis. Data points at the 80 mmHg manometric setting were excluded to prevent artificial distortion from the devices' maximum display limits.

Results

Strong linear relationships were observed between manometric and tonometric IOPs for all modes. However, the Lapine mode demonstrated proportional overestimation with steeper regression slopes compared to the Canine mode. The Canine mode showed excellent agreement (CCC: 0.967 for TVP, 0.971 for TVPR) with minimal mean bias. In contrast, the Lapine mode exhibited significant positive bias and wider limits of agreement, particularly at higher pressures.

Conclusions

Both TVP and TVPR provided accurate and reliable IOP measurements in cynomolgus monkeys when using the Canine calibration mode. Despite the central corneal thickness of cynomolgus monkeys being closer to that of rabbits, the Lapine mode resulted in significant overestimation and is not recommended. Among the tested settings, the Canine mode provided highly accurate IOP measurements with minimal bias, whereas the Lapine mode significantly overestimated IOP.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Macaca fascicularisIntraocular PressureTonometry, OcularAnimalsCalibrationReproducibility of ResultsDogs

Stichwörter der Autor:innen

cynomolgus monkeyex vivointraocular pressuremanometernon‐human primaterebound tonometer

Autor:innen und Institutionen

  • KeyPrime Research, Cheongju-si, Chungcheongbuk-do, Korea.
  • Department of Veterinary Clinical Sciences, College of Veterinary Medicine and Research Institute for Veterinary Science, Seoul National University, Seoul, Korea.
  • Eyes & Teeth Animal Clinic, Suwon, Korea.

ORCID-Profile: Wontae Kim, Seonmi Kang, Jiwon Lee, Jun Young Lee, Kangmoon Seo

Zitieren

APA

Kim, W., Kang, S., Lee, J., Lee, J. Y., & Seo, K. (2026). Ex Vivo Validation of the TONOVET Plus and TONOVET Pro in Cynomolgus Monkeys ( Macaca fascicularis ): Comparison of Canine and Lapine Calibration Modes. Veterinary Ophthalmology, 29(5). https://doi.org/10.1111/vop.70265

Vancouver

Kim W, Kang S, Lee J, Lee JY, Seo K. Ex Vivo Validation of the TONOVET Plus and TONOVET Pro in Cynomolgus Monkeys ( Macaca fascicularis ): Comparison of Canine and Lapine Calibration Modes. Veterinary Ophthalmology. 2026;29(5). doi:10.1111/vop.70265

BibTeX
@article{Kim2026,
  author = {Kim, Wontae and Kang, Seonmi and Lee, Jiwon and Lee, Jun Young and Seo, Kangmoon},
  title = {Ex Vivo Validation of the TONOVET Plus and TONOVET Pro in Cynomolgus Monkeys ( Macaca fascicularis ): Comparison of Canine and Lapine Calibration Modes},
  journal = {Veterinary Ophthalmology},
  year = {2026},
  volume = {29},
  number = {5},
  doi = {10.1111/vop.70265}
}
Kostenlos starten →