Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead
Pacing and Clinical Electrophysiology · 2026-07-20
DOI: 10.1111/pace.70369 ↗At a glance
Fallbericht in Pacing and Clinical Electrophysiology, published online on 2026-07-20. Written by Yuya Nakamura and 5 co-authors. Institution: Division of Cardiology, Department of Medicine, Showa Medical University Hospital, Tokyo, Japan. 4 references. The full text is freely available (open access, CC BY-NC 4.0).
Abstract (original language)
An 88‐year‐old pacemaker‐dependent man with a retained low‐impedance, noise‐producing right ventricular lead underwent implantation of a dual‐chamber leadless pacemaker. On postoperative day 1, he developed recurrent syncope with pauses up to 13 s. Although stored electrograms from the index events were unavailable, post‐event interrogation showed atrial pacing followed by ventricular safety pacing, and provocative testing supported persistent electrical coupling involving the retained lead. After reduction of temporary high‐output settings and reprogramming to VDD, syncope did not recur during 12 months. Retained malfunctioning leads may mediate timing‐dependent ventricular inhibition after dual‐chamber leadless pacing.
Subjects
Author keywords
Authors and institutions
- Division of Cardiology, Department of Medicine, Showa Medical University Hospital, Tokyo, Japan.
ORCID profiles: Yuya Nakamura, Takamasa Ishikawa, Shuhei Arai, Taku Asano, Toshiro Shinke
Cite
APA
Nakamura, Y., Ishikawa, T., Arai, S., Yoshikawa, K., Asano, T., & Shinke, T. (2026). Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70369
Vancouver
Nakamura Y, Ishikawa T, Arai S, Yoshikawa K, Asano T, Shinke T. Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70369
BibTeX
@article{Nakamura2026,
author = {Nakamura, Yuya and Ishikawa, Takamasa and Arai, Shuhei and Yoshikawa, Kosuke and Asano, Taku and Shinke, Toshiro},
title = {Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead},
journal = {Pacing and Clinical Electrophysiology},
year = {2026},
doi = {10.1111/pace.70369}
}
EUVIMED