EUVIMEDTutustu EUVIMEDiin
FallberichtOpen access · CC BY-NC 4.0

Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead

Yuya Nakamura iD, Takamasa Ishikawa iD, Shuhei Arai iD, Kosuke Yoshikawa, Taku Asano iD, Toshiro Shinke iD

Pacing and Clinical Electrophysiology · 2026-07-20

DOI: 10.1111/pace.70369 ↗

At a glance

Fallbericht in Pacing and Clinical Electrophysiology, published online on 2026-07-20. Written by Yuya Nakamura and 5 co-authors. Institution: Division of Cardiology, Department of Medicine, Showa Medical University Hospital, Tokyo, Japan. 4 references. The full text is freely available (open access, CC BY-NC 4.0).

TiivistelmäSubjectsBibliografiset tiedotCiteRelated studies

Abstract (original language)

An 88‐year‐old pacemaker‐dependent man with a retained low‐impedance, noise‐producing right ventricular lead underwent implantation of a dual‐chamber leadless pacemaker. On postoperative day 1, he developed recurrent syncope with pauses up to 13 s. Although stored electrograms from the index events were unavailable, post‐event interrogation showed atrial pacing followed by ventricular safety pacing, and provocative testing supported persistent electrical coupling involving the retained lead. After reduction of temporary high‐output settings and reprogramming to VDD, syncope did not recur during 12 months. Retained malfunctioning leads may mediate timing‐dependent ventricular inhibition after dual‐chamber leadless pacing.

Subjects

Author keywords

crosstalkdual‐chamber leadless pacingretained leadventricular pacing inhibitionventricular safety pacing

Authors and institutions

  • Division of Cardiology, Department of Medicine, Showa Medical University Hospital, Tokyo, Japan.

ORCID profiles: Yuya Nakamura, Takamasa Ishikawa, Shuhei Arai, Taku Asano, Toshiro Shinke

Cite

APA

Nakamura, Y., Ishikawa, T., Arai, S., Yoshikawa, K., Asano, T., & Shinke, T. (2026). Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70369

Vancouver

Nakamura Y, Ishikawa T, Arai S, Yoshikawa K, Asano T, Shinke T. Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70369