EUVIMEDEUVIMED entdecken
FallberichtOpen Access · CC BY-NC 4.0

Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead

Yuya Nakamura iD, Takamasa Ishikawa iD, Shuhei Arai iD, Kosuke Yoshikawa, Taku Asano iD, Toshiro Shinke iD

Pacing and Clinical Electrophysiology · 20.07.2026

DOI: 10.1111/pace.70369 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Pacing and Clinical Electrophysiology, online erschienen am 20.07.2026. Verfasst von Yuya Nakamura und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Medicine, Showa Medical University Hospital, Tokyo, Japan. Die Arbeit nennt 4 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

An 88‐year‐old pacemaker‐dependent man with a retained low‐impedance, noise‐producing right ventricular lead underwent implantation of a dual‐chamber leadless pacemaker. On postoperative day 1, he developed recurrent syncope with pauses up to 13 s. Although stored electrograms from the index events were unavailable, post‐event interrogation showed atrial pacing followed by ventricular safety pacing, and provocative testing supported persistent electrical coupling involving the retained lead. After reduction of temporary high‐output settings and reprogramming to VDD, syncope did not recur during 12 months. Retained malfunctioning leads may mediate timing‐dependent ventricular inhibition after dual‐chamber leadless pacing.

Schlagwörter

Stichwörter der Autor:innen

crosstalkdual‐chamber leadless pacingretained leadventricular pacing inhibitionventricular safety pacing

Autor:innen und Institutionen

  • Division of Cardiology, Department of Medicine, Showa Medical University Hospital, Tokyo, Japan.

ORCID-Profile: Yuya Nakamura, Takamasa Ishikawa, Shuhei Arai, Taku Asano, Toshiro Shinke

Zitieren

APA

Nakamura, Y., Ishikawa, T., Arai, S., Yoshikawa, K., Asano, T., & Shinke, T. (2026). Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.70369

Vancouver

Nakamura Y, Ishikawa T, Arai S, Yoshikawa K, Asano T, Shinke T. Ventricular Pacing Inhibition After Dual‐Chamber Leadless Pacing Associated With a Retained Right Ventricular Lead. Pacing and Clinical Electrophysiology. 2026. doi:10.1111/pace.70369