In Vivo Tissue Temperature Characteristics of Contact Force Catheter With a Mesh‐Shaped Irrigation Tip: A Porcine Study

Masanaru Sawada, Naoto Otsuka, Koichi Nagashima iD, Ryuta Watanabe iD, Yuji Wakamatsu, Satoshi Hayashida iD, Shu Hirata iD, Moyuru Hirata, Sayaka Kurokawa iD, Yasuo Okumura, Kenta Uto, Hiroyuki Hao und 2 weitere

Pacing and Clinical Electrophysiology · 2025

DOI: 10.1111/pace.15147 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, 2025. Verfasst von Masanaru Sawada und 13 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan. Hauptthemen laut MEDLINE: Therapeutic Irrigation, Body Temperature, Catheter Ablation, Cardiac Catheters. Die Arbeit nennt 32 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Neither the actual in vivo tissue temperatures reached with a novel contact force sensing catheter with a mesh‐shaped irrigation tip (TactiFlex SE, Abbott) nor the safety profile has been elucidated.

Methods

In a porcine model ( n = 8), thermocouples were implanted epicardially in the superior vena cava, right pulmonary vein, and esophagus close to the inferior vena cava following a right thoracotomy. After chest closure, endocardial ablation was conducted near the thermocouples under fluoroscopic guidance. We compared tissue temperatures during 50 W/13‐s high‐power short‐duration (HPSD) and 30 W/30‐s standard ablation.

Results

No steam pops were observed in 34 HPSD and 35 standard ablation applications. Tmax (maximum tissue temperature when the thermocouple was located ≤5 mm from the catheter tip) was modestly higher in HPSD compared to standard ablation (60.1°C ± 12.4°C vs. 57.8°C ± 12.9°C; p = 0.46). The peak tissue temperature correlated inversely with the catheter tip‐to‐thermocouple distance (HPSD: r = −0.40; standard: r = −0.57). Lethal temperatures (≥50°C) were reached faster with HPSD (6.5 ± 3.2 s vs. 10.3 ± 8.6 s; p = 0.04) and the distance from the catheter tip achieving a lethal tissue temperature ≥50°C (indicative of the lesion depth) was slightly shallower with HPSD (4.2 and 4.8 mm, respectively). The esophageal injury occurred superficially in both settings (0.98 ± 0.22 mm vs. 1.16 ± 0.18 mm; p = 0.29).

Conclusions

HPSD ablation produced a modestly higher and more rapid increase in the tissue temperature around the veno‐atrial junction with a shorter catheter tip‐to‐thermocouple distance required to reach lethal temperatures. This data contributes to understanding effective lesion creation and collateral injury prevention with the TactiFlex catheter.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Therapeutic IrrigationBody TemperatureCatheter AblationCardiac CathetersAnimalsSwineEquipment DesignEsophagusEquipment Failure Analysis

Stichwörter der Autor:innen

high power short duration ablationin vivo porcine tissue temperaturestandard ablation

Autor:innen und Institutionen

  • Division of Cardiology, Department of Medicine, Nihon University School of Medicine, Tokyo, Japan.
  • Department of Pathology and Microbiology, Division of Human Pathology, Nihon University School of Medicine, Tokyo, Japan.
  • Section of Laboratory for Animal Experiments, Institute of Medical Science, Medical Research Support Center, Nihon University, School of Medicine, Tokyo, Japan.

ORCID-Profile: Koichi Nagashima, Ryuta Watanabe, Satoshi Hayashida, Shu Hirata, Sayaka Kurokawa

Zitieren

APA

Sawada, M., Otsuka, N., Nagashima, K., Watanabe, R., Wakamatsu, Y., Hayashida, S., Hirata, S., Hirata, M., Kurokawa, S., Okumura, Y., Uto, K., Hao, H., Takahashi, R., & Taniguchi, Y. (2025). In Vivo Tissue Temperature Characteristics of Contact Force Catheter With a Mesh‐Shaped Irrigation Tip: A Porcine Study. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15147

Vancouver

Sawada M, Otsuka N, Nagashima K, Watanabe R, Wakamatsu Y, Hayashida S, et al. In Vivo Tissue Temperature Characteristics of Contact Force Catheter With a Mesh‐Shaped Irrigation Tip: A Porcine Study. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15147

BibTeX
@article{Sawada2025,
  author = {Sawada, Masanaru and Otsuka, Naoto and Nagashima, Koichi and Watanabe, Ryuta and Wakamatsu, Yuji and Hayashida, Satoshi and Hirata, Shu and Hirata, Moyuru and Kurokawa, Sayaka and Okumura, Yasuo and Uto, Kenta and Hao, Hiroyuki and Takahashi, Rie and Taniguchi, Yoshiki},
  title = {In Vivo Tissue Temperature Characteristics of Contact Force Catheter With a Mesh‐Shaped Irrigation Tip: A Porcine Study},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.15147}
}
Kostenlos starten →