Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardium
Europace · 31.10.2025 · 27 (11)
DOI: 10.1093/europace/euaf278 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Europace, online erschienen am 30.10.2025. Verfasst von Elio Zito und 11 weiteren Autor:innen. Beteiligte Einrichtung: Texas Cardiac Arrhythmia Institute, St David's Medical Center, 3000 N Interstate 35, Suite 700, Austin, TX 78705, USA. Hauptthemen laut MEDLINE: Therapeutic Irrigation, Cardiac Catheters, Catheter Ablation, Heart Ventricles. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Aims
Pulsed Field ablation (PFA) is a non-thermal ablation modality with functional myocardial sparing. Recent evidence suggests that clinically used PFA systems may produce non-negligible thermal effects, particularly at the tissue-catheter interface, potentially increasing the risk of thrombo-embolic complications. This study sought to characterize the temperature dynamics of two PFA pulse waveforms, delivered using a variable-loop circular catheter, under different irrigation conditions. Methods and results In total, 132 ablations were performed on 31 bovine myocardial tissue specimens, using Sequence_1 and Sequence_2 at two different irrigation rates—4 and 30 mL/min. Maximum temperatures and their rise from baseline were measured at the tissue surface, and at 3 and 7 mm depths, across different ablation conditions. Sequence_1 at 4 mL/min produced the highest surface temperature (56.4°C [54.9–58.4]) and temperature rise (ΔT: 19.4°C [17.9–21.4]). Sequence_2 at 30 mL/min showed the most favourable thermal profile, with significantly lower surface temperatures (40.8°C [37.9–43.0], ΔT: 3.8°C [0.9–6.0], P < 0.0001). At 3 mm depth, temperature increases were reduced for all settings, but remained highest with Sequence_1 at 4 mL/min (42.5°C, ΔT: 5.5°C). At 7 mm depth, temperatures remained close to baseline. Both waveform and irrigation optimization independently and synergistically reduced tissue heating, with the Sequence_2 at 30 mL/min achieving the lowest thermal load at all depths.
Conclusion
This study confirms that PFA can induce relevant thermal effects, especially at the tissue interface. However, waveform optimization and active cooling significantly mitigate these effects. Such strategies to minimize thermal effects should be implemented in clinical practice to enhance procedural safety.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Texas Cardiac Arrhythmia Institute, St David's Medical Center, 3000 N Interstate 35, Suite 700, Austin, TX 78705, USA.
- Department of Biomedical, Surgical and Dental Sciences, University of Milan, Via Festa del Perdono 7, Milan, MI 20122, Italy.
- Heart Center, Massachusetts General Hospital, Boston, MA, USA.
- Department of Cardiology, Icahn School of Medicine at Mount Sinai, One Gustave Levy Place, Box 1030, New York, NY, USA.
- Department of Biomedicine and Prevention, Division of Cardiology, University of Tor Vergata, Rome, Italy.
- Biosense Webster, Irvine, CA, USA.
ORCID-Profile: Moussa Mansour, Vivek Y Reddy, Amin Al-Ahmad, Vincenzo Mirco La Fazia, Carola Gianni, J David Burkhardt, Sanghamitra Mohanty, Luigi Di Biase
Zitieren
APA
Zito, E., Mansour, M., Reddy, V. Y., Al-Ahmad, A., La Fazia, V. M., Gianni, C., Burkhardt, J. D., Mohanty, S., Papaioannou, T., Sharma, T., Di Biase, L., & Natale, A. (2025). Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardium. Europace, 27(11). https://doi.org/10.1093/europace/euaf278
Vancouver
Zito E, Mansour M, Reddy VY, Al-Ahmad A, La Fazia VM, Gianni C, et al. Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardium. Europace. 2025;27(11). doi:10.1093/europace/euaf278
BibTeX
@article{Zito2025,
author = {Zito, Elio and Mansour, Moussa and Reddy, Vivek Y and Al-Ahmad, Amin and La Fazia, Vincenzo Mirco and Gianni, Carola and Burkhardt, J David and Mohanty, Sanghamitra and Papaioannou, Thanassis and Sharma, Tushar and Di Biase, Luigi and Natale, Andrea},
title = {Assessment of temperature dynamics in pulsed field ablation with a variable-loop circular catheter: a comparative analysis of waveform configurations and irrigation rates in specimens of bovine ventricular myocardium},
journal = {Europace},
year = {2025},
volume = {27},
number = {11},
doi = {10.1093/europace/euaf278}
}
EUVIMED