Electrophysiological Characteristics and Mechanism of His‐Bundle Pacing Guided by Electrocardiograms in Uninterrupted Real‐Time Monitoring Technique

Jinyan Zhong iD, Longfu Jiang iD, Yuelin Zhang iD, Nan Zheng, Hengdong Li

Pacing and Clinical Electrophysiology · 2025

DOI: 10.1111/pace.15106 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, 2025. Written by Jinyan Zhong and 4 co-authors. Institution: Department of Cardiology, Ningbo No. 2 Hospital, Ningbo, Zhejiang, China. Major MEDLINE topics: Bundle of His, Electrocardiography, Cardiac Pacing, Artificial. Cited 2 times so far (Crossref); 14 references. Funded by: Zhejiang Provincial Public Service and Application Research Foundation, China, Zhejiang Province Traditional Chinese Medicine Science and Technology Foundation, China, Ningbo Public Service Technology Foundation, China.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

His‐bundle pacing (HBP) is a pacing mode that provides near‐physiological pacing and has more advantages over standard right ventricle (RV) septum pacing in positive clinical results. However, traditional HBP cannot accurately and stably place the lead tip into the His‐Purkinje system. Hence, this study aimed to establish a novel strategy for HBP to strike a balance between an excellent pacing threshold and minor injuries to the conduction system.

Methods

HBP, using continuous and real‐time monitoring of unipolar His‐bundle (HB) electrograms along with electrocardiograms, was performed in 29 consecutive atrial fibrillation patients. We analyzed the unipolar HB intracardiac electrogram (EGM) and electrophysiological characteristics during HBP using John Jiang's connecting cable and an electrophysiology recording system. All patients were followed up for 12 months.

Results

Of the 29 patients, 28 (96.55%) successfully received HBP. Twenty‐one (75%) patients were confirmed to have the negative deflection of His potential with a negative amplitude of ≥ 0.1 mV in HB EGM. The subgroup with a deep negative deflection was recorded with an HBP threshold (0.71 ± 0.41 V), significantly lower than those thresholds with no deep negative deflection (1.83 ± 0.76 V) ( p < 0.05). The selective HBP rate was significantly different between the two groups: 20 (95.24%) in the deep negative deflection group and three (42.86%) in the non‐deep negative deflection group ( p < 0.05); five (17.24%) patients presented right bundle branch block (RBBB) during the lead placement.

Conclusions

This study supports the safety and feasibility of using an uninterrupted real‐time monitoring technique for HBP. An uninterrupted real‐time monitoring technique can guide the accurate placement of the HB lead and may provide a balance between an excellent pacing threshold and minor injury to the conduction system.

Subjects

MeSH terms (MEDLINE)

Bundle of HisElectrocardiographyCardiac Pacing, ArtificialHumansMaleFemaleAgedAtrial FibrillationTreatment OutcomeMiddle AgedBundle-Branch BlockComputer Systems

Author keywords

His‐bundle pacingJohn Jiang's connecting cableselective His‐bundle pacinguninterrupted real‐time monitoring technique

Authors and institutions

  • Department of Cardiology, Ningbo No. 2 Hospital, Ningbo, Zhejiang, China.

ORCID profiles: Jinyan Zhong, Longfu Jiang, Yuelin Zhang

Funding

  • Zhejiang Provincial Public Service and Application Research Foundation, China
  • Zhejiang Province Traditional Chinese Medicine Science and Technology Foundation, China
  • Ningbo Public Service Technology Foundation, China

Cite

APA

Zhong, J., Jiang, L., Zhang, Y., Zheng, N., & Li, H. (2025). Electrophysiological Characteristics and Mechanism of His‐Bundle Pacing Guided by Electrocardiograms in Uninterrupted Real‐Time Monitoring Technique. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.15106

Vancouver

Zhong J, Jiang L, Zhang Y, Zheng N, Li H. Electrophysiological Characteristics and Mechanism of His‐Bundle Pacing Guided by Electrocardiograms in Uninterrupted Real‐Time Monitoring Technique. Pacing and Clinical Electrophysiology. 2025. doi:10.1111/pace.15106

BibTeX
@article{Zhong2025,
  author = {Zhong, Jinyan and Jiang, Longfu and Zhang, Yuelin and Zheng, Nan and Li, Hengdong},
  title = {Electrophysiological Characteristics and Mechanism of His‐Bundle Pacing Guided by Electrocardiograms in Uninterrupted Real‐Time Monitoring Technique},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2025},
  doi = {10.1111/pace.15106}
}
Start for free →