EUVIMEDObjevte EUVIMED
Open access · CC BY-NC-ND 4.0

The influence of electrode‐tissue‐coverage on RF lesion formation and local impedance: Insights from an ex vivo model

Fabian Bahlke iD, Andreas Wachter, Nico Erhard, Florian Englert, Hannah Krafft, Miruna Popa, Elena Risse, Marc Kottmaier, Marta Telishevska, Sarah Lengauer, Carsten Lennerz iD, Tilko Reents and 3 more

Pacing and Clinical Electrophysiology · 2023

DOI: 10.1111/pace.14807 ↗

At a glance

Journal article in Pacing and Clinical Electrophysiology, 2023. Written by Fabian Bahlke and 14 co-authors. Institution: Department of Electrophysiology, German Heart Center Munich, Technical University of Munich (TUM), Munich, Germany. Major MEDLINE topics: Steam, Catheter Ablation. Cited 6 times so far (Crossref); 25 references. The full text is freely available (open access, CC BY-NC-ND 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Background

The influence of power, duration and contact force (CF) on radiofrequency (RF) lesion formation is well known, whereas data on local impedance (LI) and electrode‐tissue‐coverage (ETC) is scarce. The objective was to investigate their effect on lesion formation in an ex vivo model. Methods and Results An ex vivo model was developed utilizing cross‐sections of porcine heart preparations and a force‐sensing, LI‐measuring catheter. N = 72 lesion were created systematically varying ETC (minor/full), CF (1–5 g, 10–15 g, 20–25 g) and power (20 W, 30 W, 40 W, 50 W). In minor ETC, the distal tip of the catheter was in electric contact with the tissue, in full ETC the whole catheter tip was embedded within the tissue. Lesion size and all parameters were measured once per second ( n = 3320). LI correlated strongly with lesion depth ( r = −0.742 for ΔLI; r = 0.781 for %LI‐drop). Lesions in full ETC were significantly wider and deeper compared to minor ETC ( p < .001) and steam pops were more likely. Baseline LI, ΔLI, and %LI‐drop were significantly higher in full ETC ( p < .001). In lesions resulting in steam pops, baseline LI, and ΔLI were significantly higher. The influence of CF on lesion size was higher in minor ETC than in full ETC.

Conclusions

ETC is a main determinant of lesion size and occurrence of steam pops. Baseline LI and LI‐drop are useful surrogate parameters for real‐time assessment of ETC and ΔLI correlates strongly with lesion size.

Subjects

MeSH terms (MEDLINE)

SteamCatheter AblationSwineAnimalsElectric ImpedanceElectrodes

Author keywords

lesion sizelocal impedanceradiofrequency ablation

Substances

Steam

Authors and institutions

  • Department of Electrophysiology, German Heart Center Munich, Technical University of Munich (TUM), Munich, Germany.

ORCID profiles: Fabian Bahlke, Carsten Lennerz, Felix Bourier

Cite

APA

Bahlke, F., Wachter, A., Erhard, N., Englert, F., Krafft, H., Popa, M., Risse, E., Kottmaier, M., Telishevska, M., Lengauer, S., Lennerz, C., Reents, T., Hessling, G., Deisenhofer, I., & Bourier, F. (2023). The influence of electrode‐tissue‐coverage on RF lesion formation and local impedance: Insights from an ex vivo model. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14807

Vancouver

Bahlke F, Wachter A, Erhard N, Englert F, Krafft H, Popa M, et al. The influence of electrode‐tissue‐coverage on RF lesion formation and local impedance: Insights from an ex vivo model. Pacing and Clinical Electrophysiology. 2023. doi:10.1111/pace.14807

BibTeX
@article{Bahlke2023,
  author = {Bahlke, Fabian and Wachter, Andreas and Erhard, Nico and Englert, Florian and Krafft, Hannah and Popa, Miruna and Risse, Elena and Kottmaier, Marc and Telishevska, Marta and Lengauer, Sarah and Lennerz, Carsten and Reents, Tilko and Hessling, Gabriele and Deisenhofer, Isabel and Bourier, Felix},