Safety verification of a novel irrigation catheter with flexible tip of laser‐cut kerfs and contact force sensor

Junji Yamaguchi iD, Masateru Takigawa iD, Masahiko Goya iD, Tasuku Yamamoto, Takashi Ikenouchi, Hidehiro Iwakawa iD, Miho Negishi, Kentaro Goto, Takatoshi Shigeta iD, Takuro Nishimura iD, Tomomasa Takamiya, Susumu Tao iD und 2 weitere

Pacing and Clinical Electrophysiology · 2023

DOI: 10.1111/pace.14868 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Pacing and Clinical Electrophysiology, 2023. Verfasst von Junji Yamaguchi und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Tokyo Medical and Dental University Hospital, Tokyo, Japan. Hauptthemen laut MEDLINE: Steam, Catheter Ablation. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 20 Literaturangaben. Gefördert von: JSPS KAKENHI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and aims The safety evaluation of TactiFlex, a novel contact‐force sensing catheter with a flexible 4‐mm tip irrigated through laser‐cut kerfs, has been ongoing. This study aimed to verify the safety of this type of catheter.

Methods

Study 1: Radiofrequency (RF) applications at a range of powers (30–50 W), contact forces (10–20 g), and durations (10–60 s) using perpendicular/parallel catheter orientation with half‐normal (HNS) or normal saline irrigation were compared between TactiFlex (4‐mm tip) and TactiCath (3.5‐mm tip) with temperature‐controlled mode in excised porcine hearts. Study 2: The relation between RF applications using TactiFlex and the incidence of steam‐pops in the real clinical cases were examined.

Results

Study‐1: 576 RF lesions were examined. TactiFlex demonstrated a significantly lower risk of steam‐pops (5[1.7%] vs. 59[20.5%], p < .0001). Compared to 3.5‐mm‐tip catheter (TactiCath), 4‐mm‐tip catheter (TactiFlex) produced smaller lesion volume at perpendicular (193[98–554]mm 3 vs. 263[139–436]mm 3 , p < .0001), but relatively similar lesion volume at parallel contact (243[105–443]mm 3 vs. 278[180–440]mm 3 , p = .06). HNS‐irrigation tended to increase the lesion volume in both catheters and to increase the incidence of steam‐pops with TactiCath, but not with TactiFlex. The cut‐off value of %impedance‐drop ( = absolute impedance‐drop/initial impedance) of 20% predicted steam‐pops with a sensitivity = 100% and specificity = 89.6% in TactiFlex. Study‐2: 5496 RF applications in 84 patients (51AFs/8ATs/3AVNRTs/4AVRTs/17PVCs/4VTs) using TactiFlex were analyzed. Four steam‐pops (0.07%) in three patients with pericardial effusion were observed (%impedance‐drop = 24%/26%/29%/35%, respectively). The cut‐off value of %impedance‐drop = 20%, derived from ex‐vivo study, showed sensitivity = 100% and specificity = 90.1% in detecting steam‐pops.

Conclusion

TactiFlex reduced the risk of steam‐pops than TactiCath. %impedance‐drop ≤ 20% may be reasonable for safely use with a sufficient safety margin. For 4‐mm‐tip catheter, parallel‐contact may be recommended for larger lesion creation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

SteamCatheter AblationHumansAnimalsSwineTherapeutic IrrigationEquipment DesignCatheters

Stichwörter der Autor:innen

TactiFlex SEcatheter ablationlesion sizesteam-poptemperature control

Substanzen

Steam

Autor:innen und Institutionen

  • Department of Cardiovascular Medicine, Tokyo Medical and Dental University Hospital, Tokyo, Japan.
  • Department of Clinical and Diagnostic Laboratory Science, Tokyo Medical and Dental University, Tokyo, Japan.
  • Department of Cardiovascular Medicine, Akita University Graduate School of Medicine, Akita, Japan.

ORCID-Profile: Junji Yamaguchi, Masateru Takigawa, Masahiko Goya, Hidehiro Iwakawa, Takatoshi Shigeta, Takuro Nishimura, Susumu Tao

Förderung

  • JSPS KAKENHI

Zitieren

APA

Yamaguchi, J., Takigawa, M., Goya, M., Yamamoto, T., Ikenouchi, T., Iwakawa, H., Negishi, M., Goto, K., Shigeta, T., Nishimura, T., Takamiya, T., Tao, S., Miyazaki, S., & Sasano, T. (2023). Safety verification of a novel irrigation catheter with flexible tip of laser‐cut kerfs and contact force sensor. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14868

Vancouver

Yamaguchi J, Takigawa M, Goya M, Yamamoto T, Ikenouchi T, Iwakawa H, et al. Safety verification of a novel irrigation catheter with flexible tip of laser‐cut kerfs and contact force sensor. Pacing and Clinical Electrophysiology. 2023. doi:10.1111/pace.14868

BibTeX
@article{Yamaguchi2023,
  author = {Yamaguchi, Junji and Takigawa, Masateru and Goya, Masahiko and Yamamoto, Tasuku and Ikenouchi, Takashi and Iwakawa, Hidehiro and Negishi, Miho and Goto, Kentaro and Shigeta, Takatoshi and Nishimura, Takuro and Takamiya, Tomomasa and Tao, Susumu and Miyazaki, Shinsuke and Sasano, Tetsuo},
  title = {Safety verification of a novel irrigation catheter with flexible tip of laser‐cut kerfs and contact force sensor},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2023},
  doi = {10.1111/pace.14868}
}
Kostenlos starten →