Averaging real-time impedance enhances the prediction of steam pop risk and lesion characteristics
Europace · 2025
DOI: 10.1093/europace/euaf315 ↗Kurz gefasst
Fachartikel in der Zeitschrift Europace, 2025. Verfasst von Hidehiro Iwakawa und 18 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiovascular Medicine, Institute of Science Tokyo, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan. Hauptthemen laut MEDLINE: Steam, Catheter Ablation, Heart Ventricles, Heart Atria. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Japan Society for the Promotion of Science.
Abstract (Originalsprache)
Aims
A novel impedance filtering function that averages impedance values was developed to mitigate cardiac and respiratory oscillations. We aimed to evaluate the clinical significance of averaging real-time impedance in predicting steam pops (SPs) and lesion characteristics. Methods and results Radiofrequency (RF) ablation was performed in 20 swine using a flexible-tip temperature-controlled power regulation catheter. Both unfiltered and filtered (averaged) impedance values were recorded using the EnSite™ X system. For each RF application, absolute (ΔImp-drop) and relative (%Imp-drop) impedance drops were quantified. Associations between impedance parameters and SP occurrence, atrial lesion transmurality, and ventricular lesion dimensions were evaluated. Among 959 lesions, SPs occurred in 36 applications (3.8%), all within the ventricles. Notably, 6 SPs occurred within 90 s despite RF power ≤ 40 W, with 4 during left ventricular ablation under low systolic blood pressure (<40 mmHg). Lesions with SPs exhibited significantly greater unfiltered and averaged ΔImp-drop and %Imp-drop (all P < 0.001). Averaged %Imp-drop showed the highest predictive value for SPs (AUC = 0.93), with a 20.9% cut-off yielding 88.9% sensitivity and 85.5% specificity. The time to reach the initial 10%, 15%, and 20% reduction in averaged %Imp-drop was not associated with SP occurrence. Both unfiltered and averaged impedance drops correlated with atrial transmural lesion formation. Averaged impedance drops significantly improved estimation of lesion depth, surface area, and volume compared with unfiltered values (P < 0.01).
Conclusion
The averaged relative impedance drop demonstrated the strongest association with SP occurrence, and averaging impedance provided a more accurate assessment of lesion characteristics than unfiltered measurements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Cardiovascular Medicine, Institute of Science Tokyo, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
- Department of Cardiovascular Medicine, Akita University Graduate School of Medicine, Akita, Japan.
- Division of Advanced Arrhythmia Research, Institute of Science Tokyo, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
- Department of Clinical and Diagnostic Laboratory Science, Institute of Science Tokyo, Tokyo, Japan.
- Japan Small Animal Medical Center, Saitama, Japan.
- Animal Research Facilities, Bioscience Center, Research Infrastructure Management Center, Institute of Science Tokyo, Tokyo, Japan.
ORCID-Profile: Hidehiro Iwakawa, Masateru Takigawa, Junji Yamaguchi, Ryosuke Kato, Masaki Honda, Ryo Tateishi, Miho Negishi, Iwanari Kawamura, Kentaro Goto, Kensuke Ihara, Takuro Nishimura, Kazuya Yamao, Susumu Tao, Sayaka Suzuki, Iichiro Onishi, Shinsuke Miyazaki, Hiroyuki Watanabe, Tetsuo Sasano
Förderung
- Japan Society for the Promotion of Science
Zitieren
APA
Iwakawa, H., Takigawa, M., Yamaguchi, J., Kato, R., Honda, M., Tateishi, R., Negishi, M., Kawamura, I., Goto, K., Ihara, K., Nishimura, T., Yamao, K., Tao, S., Suzuki, S., Iwanaga, T., Onishi, I., Miyazaki, S., Watanabe, H., & Sasano, T. (2025). Averaging real-time impedance enhances the prediction of steam pop risk and lesion characteristics. Europace. https://doi.org/10.1093/europace/euaf315
Vancouver
Iwakawa H, Takigawa M, Yamaguchi J, Kato R, Honda M, Tateishi R, et al. Averaging real-time impedance enhances the prediction of steam pop risk and lesion characteristics. Europace. 2025. doi:10.1093/europace/euaf315
BibTeX
@article{Iwakawa2025,
author = {Iwakawa, Hidehiro and Takigawa, Masateru and Yamaguchi, Junji and Kato, Ryosuke and Honda, Masaki and Tateishi, Ryo and Negishi, Miho and Kawamura, Iwanari and Goto, Kentaro and Ihara, Kensuke and Nishimura, Takuro and Yamao, Kazuya and Tao, Susumu and Suzuki, Sayaka and Iwanaga, Takehiro and Onishi, Iichiro and Miyazaki, Shinsuke and Watanabe, Hiroyuki and Sasano, Tetsuo},
title = {Averaging real-time impedance enhances the prediction of steam pop risk and lesion characteristics},
journal = {Europace},
year = {2025},
doi = {10.1093/europace/euaf315}
}
EUVIMED