EUVIMEDObjevte EUVIMED
FallberichtOpen access · CC BY-NC-ND 4.0

Percutaneous extraction of a Micra AV transcatheter pacing system due to a rare sudden battery failure after 19 months from implantation: A first experience worldwide

Vincenzo Ezio Santobuono iD, Paolo Basile, Maria Cristina Carella, Francesca Pomarico, Stefano Favale, Marco Matteo Ciccone, Andrea Igoren Guaricci iD

Pacing and Clinical Electrophysiology · 2024

DOI: 10.1111/pace.14714 ↗

At a glance

Fallbericht in Pacing and Clinical Electrophysiology, 2024. Written by Vincenzo Ezio Santobuono and 6 co-authors. Institution: Interdisciplinary Department of Medicine, University Cardiology Unit, Policlinic University Hospital, Bari, Italy. Major MEDLINE topics: Arrhythmias, Cardiac, Pacemaker, Artificial. Cited 3 times so far (Crossref); 15 references. The full text is freely available (open access, CC BY-NC-ND 4.0).

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Introduction

Micra AV Transcatheter Pacing System (TPS) represents an innovative second‐generation leadless pacemaker which represents an effective alternative to conventional devices in selected cases. Intrinsic malfunctions of these devices are rare, requiring sometimes their retrieval. When performed in experienced centers, this procedure is safe. Case Presentation We describe a case of sudden battery malfunction of a Micra AV TPS, which required the extraction and the placement of a new pacing system in the right ventricle.

Discussion

This case, which has never been reported, highlights the need to a careful fluoroscopic evaluation and the usefulness of remote monitoring.

Subjects

MeSH terms (MEDLINE)

Arrhythmias, CardiacPacemaker, ArtificialHumansTreatment OutcomeEquipment DesignCardiac Pacing, Artificial

Author keywords

Micra AVbattery malfunctioncardiac pacingextractionfluoroscopyleadless pacemakerremote monitoring

Authors and institutions

  • Interdisciplinary Department of Medicine, University Cardiology Unit, Policlinic University Hospital, Bari, Italy.
  • Department of Precision and Regenerative medicine and Ionian area, Policlinic University Hospital, Bari, Italy.

ORCID profiles: Vincenzo Ezio Santobuono, Andrea Igoren Guaricci

Cite

APA

Santobuono, V. E., Basile, P., Carella, M. C., Pomarico, F., Favale, S., Ciccone, M. M., & Guaricci, A. I. (2024). Percutaneous extraction of a Micra AV transcatheter pacing system due to a rare sudden battery failure after 19 months from implantation: A first experience worldwide. Pacing and Clinical Electrophysiology. https://doi.org/10.1111/pace.14714

Vancouver

Santobuono VE, Basile P, Carella MC, Pomarico F, Favale S, Ciccone MM, et al. Percutaneous extraction of a Micra AV transcatheter pacing system due to a rare sudden battery failure after 19 months from implantation: A first experience worldwide. Pacing and Clinical Electrophysiology. 2024. doi:10.1111/pace.14714

BibTeX
@article{Santobuono2024,
  author = {Santobuono, Vincenzo Ezio and Basile, Paolo and Carella, Maria Cristina and Pomarico, Francesca and Favale, Stefano and Ciccone, Marco Matteo and Guaricci, Andrea Igoren},
  title = {Percutaneous extraction of a Micra AV transcatheter pacing system due to a rare sudden battery failure after 19 months from implantation: A first experience worldwide},
  journal = {Pacing and Clinical Electrophysiology},
  year = {2024},
  doi = {10.1111/pace.14714}
}
Start for free →