EUVIMEDEUVIMED entdecken
Multizentrische Studie

Multicenter Results of a Novel Pediatric Pacemaker in Neonates and Infants

Charles I Berul iD, Lindsey Haack iD, Elizabeth D Sherwin iD, Robert D Whitehill iD, Dustin Nash, Soham Dasgupta, Christopher Johnsrude iD, Stephanie F Chandler iD, Marc D LeGras iD, Bradley C Clark iD, Erick Jimenez iD, Heather M Giacone iD und 9 weitere

Circulation: Arrhythmia and Electrophysiology · 2025-03 · 18 (3)

DOI: 10.1161/circep.124.013436 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Circulation: Arrhythmia and Electrophysiology, 2025-03. Verfasst von Charles I Berul und 20 weiteren Autor:innen. Beteiligte Einrichtung: Children's National Hospital, Washington, DC (C.I.B., L.H., E.D.S.). Hauptthemen laut MEDLINE: Arrhythmias, Cardiac, Cardiac Pacing, Artificial, Heart Defects, Congenital, Heart Rate. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 13 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: To address the unmet need for a smaller pacemaker for babies, a specially modified implantable pulse generator was developed containing a Medtronic Micra subassembly in a polymer header connecting to a bipolar epicardial lead. The aim of this study was to report midterm follow-up data and outcomes of patients who underwent implantation of this device. METHODS: Deidentified data were collected from 12 of 15 sites in the United States implanting the pediatric implantable pulse generator between March 2022 and February 2024. All 29 patients at these 12 sites within this timeframe were included in the analysis. RESULTS: The median age at implant was 15 days (range, 0 days to 3 years, including 1 outlier). The median weight was 2.3 kg (range, 1.3–11.4 kg). Gestational age was 28.5 weeks to term, with 23 (79%) patients born prematurely. Of those with anatomic information, 25% had congenital heart disease. The average duration of implant was 325 days (73–808 days). The most recent lead impedance mean was 612 ohms (450–840 ohms), ventricular capture threshold mean was 1 V @ 0.4 ms (range, 0.38–2.75 V), and R-wave sensing mean was 12.5 mV (3.6–20 mV). There were 7 generator explants (24%), removed at 6.5 to 31 months of age. CONCLUSIONS: The pediatric implantable pulse generator can be safely implanted in neonates and infants. This multicenter report demonstrates that the devices remain stable, with effective pacing, normal electrical parameters, and battery longevity aligned with projections. This novel pediatric pacemaker provides a viable alternative to standard-size generators and addresses a vital unmet need for these small patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arrhythmias, CardiacCardiac Pacing, ArtificialHeart Defects, CongenitalHeart RatePacemaker, ArtificialChild, PreschoolFemaleHumansInfantInfant, NewbornMaleAge FactorsEquipment DesignProsthesis DesignTime FactorsTreatment OutcomeUnited StatesFollow-Up Studies

Stichwörter der Autor:innen

childelectric impedancegestational ageinfant, newbornpacemaker, artificial

Autor:innen und Institutionen

  • Children's National Hospital, Washington, DC (C.I.B., L.H., E.D.S.).
  • Children's Healthcare of Atlanta, GA (R.D.W.).
  • Children's Hospital of Colorado, Denver (D.N.).
  • Norton Children's Hospital, Louisville, KY (S.D., C.J.).
  • Lurie Children's Hospital of Chicago, IL (S.F.C.).
  • Pediatric Cardiology Center of Oregon, Portland (M.D.L.G.).

ORCID-Profile: Charles I Berul, Lindsey Haack, Elizabeth D Sherwin, Robert D Whitehill, Christopher Johnsrude, Stephanie F Chandler, Marc D LeGras, Bradley C Clark, Erick Jimenez, Heather M Giacone, Scott Ceresnak, Santiago O Valdes, Edward Rhee, Rachel Brucker, Sonja Olson

Zitieren

APA

Berul, C. I., Haack, L., Sherwin, E. D., Whitehill, R. D., Nash, D., Dasgupta, S., Johnsrude, C., Chandler, S. F., LeGras, M. D., Clark, B. C., Jimenez, E., Giacone, H. M., Ceresnak, S., Goya, G., Jackson, L. B., Pham, T. D., Valdes, S. O., Rhee, E., Brucker, R., …, & Whitman, T. (2025). Multicenter Results of a Novel Pediatric Pacemaker in Neonates and Infants. Circulation: Arrhythmia and Electrophysiology, 18(3). https://doi.org/10.1161/circep.124.013436

Vancouver

Berul CI, Haack L, Sherwin ED, Whitehill RD, Nash D, Dasgupta S, et al. Multicenter Results of a Novel Pediatric Pacemaker in Neonates and Infants. Circulation: Arrhythmia and Electrophysiology. 2025;18(3). doi:10.1161/circep.124.013436

BibTeX
@article{Berul2025,
  author = {Berul, Charles I and Haack, Lindsey and Sherwin, Elizabeth D and Whitehill, Robert D and Nash, Dustin and Dasgupta, Soham and Johnsrude, Christopher and Chandler, Stephanie F and LeGras, Marc D and Clark, Bradley C and Jimenez, Erick and Giacone, Heather M and Ceresnak, Scott and Goya, Garbiñe and Jackson, Lanier B and Pham, Tam Dan and Valdes, Santiago O and Rhee, Edward and Brucker, Rachel and Olson, Sonja and Whitman, Teri},
  title = {Multicenter Results of a Novel Pediatric Pacemaker in Neonates and Infants},
  journal = {Circulation: Arrhythmia and Electrophysiology},
  year = {2025},
  volume = {18},
  number = {3},
  doi = {10.1161/circep.124.013436}
}