Multicenter Study of the Electrical Characteristics and Short‐Term Outcomes of the Aveir VR Leadless Pacemaker
Journal of the American Heart Association · 2026
DOI: 10.1161/jaha.126.050579 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of the American Heart Association, 2026. Verfasst von Jian-du Yang und 16 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Cardiovascular Disease, Arrhythmia Center, Fuwai Hospital, National Center for Cardiovascular Diseases Chinese Academy of Med…. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Cardiac Pacing, Artificial, Arrhythmias, Cardiac. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Background
The Aveir leadless pacemaker uses an active fixation method, enabling real‐time monitoring of electrical parameters during implantation. However, comprehensive studies on the electrical parameters during this procedure are rare. This study aims to analyze the electrical characteristics to further guide the implantation strategy and improve device stability and safety.
Methods
This multicenter retrospective study enrolled 119 patients (mean age, 70.18 years; 59.58% women) who received the Aveir VR leadless pacemaker from November 2024 to May 2025 at 10 centers in China. Intraprocedural variations in commanded electrogram, current of injury (COI), impedance, pacing threshold, and sensing parameters were meticulously documented.
Results
Commanded electrogram mapping demonstrated various morphologies (R, RS, QR, QRS, and QS), aiding localization. During fixation, 58.82% of patients exhibited an increased COI from mapping to 0.5 turns, which was associated with reduced short‐term pacing thresholds. From 0.5 to 1 turn, 52.94% showed a further increase in COI. Receiver operating characteristic analysis revealed that an impedance increase has predictive value for short‐term pacing thresholds, with an area under the receiver operating characteristic curve of 0.634 and a cutoff value of 230 Ω (sensitivity, 0.622; specificity, 0.41). Lead stability showed a moderate correlation with impedance increase ( R =0.44, P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- State Key Laboratory of Cardiovascular Disease, Arrhythmia Center, Fuwai Hospital, National Center for Cardiovascular Diseases Chinese Academy of Medical Sciences and Peking Union Medical College Beijing China.
- Department of Cardiology, Shanghai Chest Hospital Shanghai Jiao Tong University Shanghai China.
- Department of Cardiology West China Hospital of Sichuan University Chengdu China.
- Department of Cardiology, the First Affiliated Hospital Xi'an Jiaotong University Xi'an China.
- Department of Cardiology First Affiliated Hospital of Xinjiang Medical University Xinjiang China.
- Department of Cardiology The Second People's Hospital of Chengdu Chengdu Sichuan China.
ORCID-Profile: Jian-du Yang, Ruo-Gu Li, Jiang-Hua Zhang, Bao-Jian Zhang, Hui Luo, Zhong-Li Chen, Baopeng Tang, Ke-Ping Chen
Zitieren
APA
Yang, J. D., Li, R. G., Liu, X. B., Xue, X. L., Zhang, J. H., Hu, Y. M., Zhang, B. J., Tong, L., Luo, H., Shen, M., Chen, Z. L., Aiyasiding, X., Cai, M. X., Chi, X., Dai, Y., Tang, B., & Chen, K. P. (2026). Multicenter Study of the Electrical Characteristics and Short‐Term Outcomes of the Aveir VR Leadless Pacemaker. Journal of the American Heart Association. https://doi.org/10.1161/jaha.126.050579
Vancouver
Yang JD, Li RG, Liu XB, Xue XL, Zhang JH, Hu YM, et al. Multicenter Study of the Electrical Characteristics and Short‐Term Outcomes of the Aveir VR Leadless Pacemaker. Journal of the American Heart Association. 2026. doi:10.1161/jaha.126.050579
BibTeX
@article{Yang2026,
author = {Yang, Jian-du and Li, Ruo-Gu and Liu, Xing-Bin and Xue, Xiao-Lin and Zhang, Jiang-Hua and Hu, Yong-Mei and Zhang, Bao-Jian and Tong, Lin and Luo, Hui and Shen, Min and Chen, Zhong-Li and Aiyasiding, Xiahenazi and Cai, Meng-Xing and Chi, Xiaojian and Dai, Yan and Tang, Baopeng and Chen, Ke-Ping},
title = {Multicenter Study of the Electrical Characteristics and Short‐Term Outcomes of the Aveir VR Leadless Pacemaker},
journal = {Journal of the American Heart Association},
year = {2026},
doi = {10.1161/jaha.126.050579}
}
EUVIMED