Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort
Europace · 2025
DOI: 10.1093/europace/euaf291 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Europace, 2025. Verfasst von Peggy Jacon und 31 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, University Hospital of Grenoble Alpes, 38043 Grenoble, France. Hauptthemen laut MEDLINE: Pacemaker, Artificial, Device Removal, Reoperation, Arrhythmias, Cardiac. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Aims
Leadless pacemakers (LLPs) offer a valuable option for selected patients requiring ventricular pacing. However, data on revision procedures—defined as any intervention on an implanted LLP to correct dysfunction or upgrade the system, with or without removal—remain scarce. This study aimed to evaluate the real-world incidence, characteristics, and outcomes of LLP revisions, with particular emphasis on the feasibility and safety of removal. Methods and results This multicentre retrospective study involved 46 French centres performing LLP implantations between 2015 and 2023. Among 8994 Micra LLPs implanted, 100 revision procedures (1.1%) were performed in 100 patients, including 73 Micra VR and 27 Micra AV devices, at a median follow-up of 229 (68.5–629.8) days. Seventy revisions were managed without device removal, while 30 involved removal. Revision indications included device upgrade (55%), high pacing threshold (33%), battery depletion (3%), ventricular arrhythmias (3%), and miscellaneous causes (6%). No significant differences were observed in pacing indications, implantation characteristics, or initial device indication between the two groups. Revisions occurred significantly earlier in the removal group than in the group without removal [median 81.5 (8–211.5) days, vs. 334 (130.25–882.5) days, P < 0.001], with 80% occurring within 1 year. Device removal was successful in 97% of cases (one failure at 157 days of follow-up), with no major complications reported.
Conclusion
LLP revisions are infrequent and predominantly managed without device removal. However, given the high procedural success rate and the low complication rate, device removal appears to be underutilized and should be considered more frequently on a case-by-case basis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, University Hospital of Grenoble Alpes, 38043 Grenoble, France.
- Department of Cardiology, University Hospital of Angers, Angers, France.
- Department of Cardiology, University Hospital of Tours, Tours, France.
- Department of Cardiology, European Georges Pompidou Hospital, Paris, France.
- Department of Cardiology, University Hospital Saint Etienne, Saint Etienne, France.
- Department of Cardiology, University Hospital of Lyon, Lyon, France.
ORCID-Profile: Peggy Jacon, Mouna Benkilani, Arnaud Bisson, Eloi Marijon, Antoine Da Costa, Christelle Haddad, Nathalie Behar, Olivier Cesari, Nicolas Lellouche, Vincent Mansourati, Jean Luc Pasquié, Romain Eschalier, Pierre Ollitrault, Hugues Blangy, Pierre Mondoly, Xavier Waintraub, Antoine Andorin, Frederic Anselme, Pierre Frey, Rodrigue Garcia, Benoit Guy-Moyat, Christelle Marquié, Alexis Mechulan, Baptiste Maille, Nathan Vaxelaire, Adrien Carabelli, Sandrine Venier, Pascal Defaye
Zitieren
APA
Jacon, P., Benkilani, M., Bisson, A., Marijon, E., Da Costa, A., Haddad, C., Behar, N., Cesari, O., Lellouche, N., Mansourati, V., Pasquié, J. L., Eschalier, R., Ollitrault, P., Blangy, H., Bordachar, P., Jesel Morel, L., Mondoly, P., Waintraub, X., Andorin, A., …, & Defaye, P. (2025). Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort. Europace. https://doi.org/10.1093/europace/euaf291
Vancouver
Jacon P, Benkilani M, Bisson A, Marijon E, Da Costa A, Haddad C, et al. Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort. Europace. 2025. doi:10.1093/europace/euaf291
BibTeX
@article{Jacon2025,
author = {Jacon, Peggy and Benkilani, Mouna and Bisson, Arnaud and Marijon, Eloi and Da Costa, Antoine and Haddad, Christelle and Behar, Nathalie and Cesari, Olivier and Lellouche, Nicolas and Mansourati, Vincent and Pasquié, Jean Luc and Eschalier, Romain and Ollitrault, Pierre and Blangy, Hugues and Bordachar, Pierre and Jesel Morel, Laurence and Mondoly, Pierre and Waintraub, Xavier and Andorin, Antoine and Anselme, Frederic and Frey, Pierre and Garcia, Rodrigue and Guy-Moyat, Benoit and Serzian, Guillaume and Marquié, Christelle and Mechulan, Alexis and Maille, Baptiste and Vaxelaire, Nathan and Benali, Rémi and Carabelli, Adrien and Venier, Sandrine and Defaye, Pascal},
title = {Micra leadless pacemaker revisions: incidence, characteristics, and outcomes from a multicentre French cohort},
journal = {Europace},
year = {2025},
doi = {10.1093/europace/euaf291}
}
EUVIMED