Effect of implant shape and length on the accuracy of robot‐assisted immediate implant surgery: An in vitro study
Clinical Oral Implants Research · 2024
DOI: 10.1111/clr.14232 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Oral Implants Research, 2024. Verfasst von Yunxiao Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Oral and Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key L…. Hauptthemen laut MEDLINE: Dental Implants, Robotic Surgical Procedures, Robotics, Surgery, Computer-Assisted. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 27 Literaturangaben. Gefördert von: Fundamental Research Funds for the Central Universities.
Abstract (Originalsprache)
Objectives
To compare the accuracy of immediate implant placement of cylindrical implants (CI) and tapered implants (TI) of different lengths using a robotic dental implant system. Materials and Methods CI and TI of three lengths (8, 10, and 12 mm) each were digitally planned and placed in a three‐dimensional printed extraction socket model under robotic guidance. There were six groups with three samples in each group, resulting in a total of 18 samples. Implant angular deviation, platform point deviation (total, lateral, depth), and implant apical point deviation (total, lateral, depth) were recorded and compared between the different groups.
Results
The angular deviations for CI 8 mm, CI 10 mm, CI 12 mm, TI 8 mm, TI 10 mm, and TI 12 mm were 1.32° ± 0.19°, 1.03° ± 0.56°, 1.31° ± 0.38°, 1.27° ± 0.64°, 1.10° ± 0.43° and 1.05° ± 0.45°, respectively. The total deviations of platform and apical points for CI 8 mm, CI 10 mm, CI 12 mm, TI 8 mm, TI 10 mm, and TI 12 mm were 0.79 ± 0.18 mm, 0.77 ± 0.33 mm; 0.64 ± 0.21 mm, 0.55 ± 0.17 mm; 0.64 ± 0.37 mm, 0.65 ± 0.34 mm; 0.68 ± 0.26 mm, 0.71 ± 0.20 mm; 0.70 ± 0.12 mm, 0.66 ± 0.23 mm; and 0.71 ± 0.15 mm, 0.77 ± 0.29 mm, respectively, and had no significant differences.
Conclusions
Within the limitation of this study, acceptable accuracy can be achieved for both TI and CI using robotic systems. Our study demonstrated that the implant shape and length did not affect the accuracy of immediate implant placement under robotic guidance in vitro. However, further trials are required to confirm their efficacy in clinical practice.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- State Key Laboratory of Oral and Maxillofacial Reconstruction and Regeneration, Key Laboratory of Oral Biomedicine Ministry of Education, Hubei Key Laboratory of Stomatology, School and Hospital of Stomatology, Wuhan Un…
- Department of Oral Implantology, School and Hospital of Stomatology, Wuhan University, Wuhan, China.
- Oral Biomaterials and Application Technology Engineering Research Center of Hubei Province, Wuhan, China.
ORCID-Profile: Yunxiao Wang, Shimin Yu, Yulan Wang, Yi Feng, Qi Yan, Yufeng Zhang
Förderung
- Fundamental Research Funds for the Central Universities
Zitieren
APA
Wang, Y., Yu, S., Wang, Y., Feng, Y., Yan, Q., & Zhang, Y. (2024). Effect of implant shape and length on the accuracy of robot‐assisted immediate implant surgery: An in vitro study. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14232
Vancouver
Wang Y, Yu S, Wang Y, Feng Y, Yan Q, Zhang Y. Effect of implant shape and length on the accuracy of robot‐assisted immediate implant surgery: An in vitro study. Clinical Oral Implants Research. 2024. doi:10.1111/clr.14232
BibTeX
@article{Wang2024,
author = {Wang, Yunxiao and Yu, Shimin and Wang, Yulan and Feng, Yi and Yan, Qi and Zhang, Yufeng},
title = {Effect of implant shape and length on the accuracy of robot‐assisted immediate implant surgery: An in vitro study},
journal = {Clinical Oral Implants Research},
year = {2024},
doi = {10.1111/clr.14232}
}
EUVIMED