Accuracy of dental implant placement with a robotic system in partially edentulous patients: A prospective, single‐arm clinical trial
Clinical Oral Implants Research · 2023
DOI: 10.1111/clr.14083 ↗Kurz gefasst
Klinische Studie in der Zeitschrift Clinical Oral Implants Research, 2023. Verfasst von Wei Chen und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Implantology, Affiliated Stomatological Hospital of Nanjing Medical University, Nanjing, China. Hauptthemen laut MEDLINE: Dental Implants, Robotic Surgical Procedures, Mouth, Edentulous, Surgery, Computer-Assisted. Die Arbeit wurde bisher 76-mal zitiert (Crossref) und nennt 52 Literaturangaben. Gefördert von: International Science and Technology Cooperation Program of China, National Natural Science Foundation of China, National Science Fund for Distinguished Young Scholars of China.
Abstract (Originalsprache)
Objectives
This clinical study aimed to assess the accuracy of implant positions using a robotic system in partially edentulous patients. Materials and Methods Twenty‐eight partially edentulous patients received 31 implants using the robotic system. Deviations between the planned and placed implants were calculated after surgery. The deviations were compared with objective performance goals (OPGs) from reported studies of fully guided static computer‐assisted implant surgery (CAIS) and dynamic CAIS. A multiple linear regression analysis was performed to investigate the possible effects of the type and side of the arch, implant location, and implant dimensions on the deviations.
Results
The evaluation of 31 implants resulted in a mean angle deviation of 2.81 ± 1.13° (95% confidence interval (CI): 2.40–3.23°), while the 3D deviations at the implant shoulder and apex were 0.53 ± 0.23 mm (95% CI 0.45–0.62 mm) and 0.53 ± 0.24 mm (95% CI 0.44–0.61 mm), respectively. The upper limits of the 95% CI of 3D deviations were lower than those of the corresponding OPGs; however, the angle deviation was similar to that of the OPG. No statistically significant differences were found for the type and side of the arch, implant location, and implant dimensions to the deviations ( p > .05).
Conclusions
The robotic system appears to achieve higher accuracy in implant positions than static and dynamic CAIS in partially edentulous patients (Chinese Clinical Trial Registry ChiCTR2300067587).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Oral Implantology, Affiliated Stomatological Hospital of Nanjing Medical University, Nanjing, China.
- Jiangsu Province Key Laboratory of Oral Diseases, Nanjing, China.
- Department of Orthodontic, Affiliated Stomatological Hospital of Nanjing Medical University, Nanjing, China.
- Jiangsu Province Engineering Research Center of Stomatological Translational Medicine, Nanjing, China.
- Hangzhou Lancet Robotics Company Ltd, Hangzhou, China.
ORCID-Profile: Chunbo Tang
Förderung
- International Science and Technology Cooperation Program of China
- National Natural Science Foundation of China
- National Science Fund for Distinguished Young Scholars of China
- Southeast University and Nanjing Medical University Cooperative Research Project
Zitieren
APA
Chen, W., Al-Taezi, K. A., Chu, C. H., Shen, Y., Wu, J., Cai, K., Chen, P., & Tang, C. (2023). Accuracy of dental implant placement with a robotic system in partially edentulous patients: A prospective, single‐arm clinical trial. Clinical Oral Implants Research. https://doi.org/10.1111/clr.14083
Vancouver
Chen W, Al-Taezi KA, Chu CH, Shen Y, Wu J, Cai K, et al. Accuracy of dental implant placement with a robotic system in partially edentulous patients: A prospective, single‐arm clinical trial. Clinical Oral Implants Research. 2023. doi:10.1111/clr.14083
BibTeX
@article{Chen2023,
author = {Chen, Wei and Al-Taezi, Khulood Ali and Chu, Catherine Huihan and Shen, Yue and Wu, Jin and Cai, Kunzhan and Chen, Peng and Tang, Chunbo},
title = {Accuracy of dental implant placement with a robotic system in partially edentulous patients: A prospective, single‐arm clinical trial},
journal = {Clinical Oral Implants Research},
year = {2023},
doi = {10.1111/clr.14083}
}
EUVIMED