Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts
IEEE Transactions on Radiation and Plasma Medical Sciences · 2026-05 · 10 (5) · 753-763
DOI: 10.1109/trpms.2025.3624609 ↗Kurz gefasst
Fachartikel in der Zeitschrift IEEE Transactions on Radiation and Plasma Medical Sciences, 2026-05. Verfasst von Lukas U Rauscher und 8 weiteren Autor:innen. Beteiligte Einrichtung: Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, 72076 Tuebingen, Germany. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIBIB NIH HHS, Deutsche Forschungsgemeinschaft, BRAIN Initiative NIH-NIBIB & NINDS.
Abstract (Originalsprache)
High-sensitivity organ-specific PET inserts employing long scintillation crystals in combination with small ring diameters require depth-of-interaction (DOI) encoding to ensure uniform spatial resolution across the field of view. In this study, we have developed high-resolution, DOI-capable detectors based on the Hamamatsu C13500 TOF-PET modules, using 1.6W1.6 mm2 LSO crystals coupled to 4W4 mm2 SiPMs. Detector configurations were designed either for a brain PET/MR insert (26 mm long crystals with a top light guide (TLG) for DOI encoding via light spread) or for a breast PET/MR insert (15 mm long with a TLG; 15 mm long with a stacked dual-layer configuration). Corresponding reference configurations without DOI capability were included to assess the impact of DOI implementation on detector performance. The 26 mm long brain detector configuration with TLG achieved an energy resolution of 12.1% and a coincidence time resolution (CTR) of 684 ps, with a degradation of 148 ps compared to the non-DOI variant. A newly developed DOI parameter calculation method improved DOI resolution from 9.1 1 1.7 mm to 7.1 1 1.8 mm. The 15 mm long single layer breast detector configuration with TLG yielded an energy resolution of 10.8%, a CTR of 460 ps, and a DOI resolution of 6.3 1 1.8 mm. The 15 mm dual layer detector configuration provided comparable performance (11.5% energy resolution, 605 ps CTR) while eliminating the need for DOI calibration. For both the 15 and 26 mm single layer crystal block configurations, introducing a TLG preserved crystal identification and energy resolution. Overall, the DOI-capable designs demonstrated only minor performance trade-offs, supporting the feasibility of both approaches for integration into organ-specific PET/MR inserts.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, 72076 Tuebingen, Germany.
- Department of Physics, Massachusetts Institute of Technology, Cambridge, MA 02139, USA and Athinoula A. Martinos Center for Biomedical Imaging, Radiology Department, Massachusetts General Hospital, Boston, MA 02129, USA.
- Molecular Imaging, Siemens Medical Solutions USA Inc., Knoxville, USA.
- Athinoula A. Martinos Center for Biomedical Imaging, Radiology Department, Massachusetts General Hospital and Harvard Medical School, Boston, MA 02129, USA.
- Nuclear Physics Group and IPARCOS, Universidad Complutense de Madrid, Madrid, Spain.
- Werner Siemens Imaging Center, Department of Preclinical Imaging and Radiopharmacy, 72076 Tuebingen, Germany and also with Cluster of Excellence iFIT (EXC 2180) "Image Guided and Functionally Instructed Tumor Therapies"…
Förderung
- NIBIB NIH HHS
- Deutsche Forschungsgemeinschaft
- BRAIN Initiative NIH-NIBIB & NINDS
Zitieren
APA
Rauscher, L. U., Allen, M. S., Judenhofer, M. S., Byars, L., Scipioni, M., Udias, J. M., Catana, C., Pichler, B. J., & Schmidt, F. P. (2026). Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts. IEEE Transactions on Radiation and Plasma Medical Sciences, 10(5), 753-763. https://doi.org/10.1109/trpms.2025.3624609
Vancouver
Rauscher LU, Allen MS, Judenhofer MS, Byars L, Scipioni M, Udias JM, et al. Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts. IEEE Transactions on Radiation and Plasma Medical Sciences. 2026;10(5):753-763. doi:10.1109/trpms.2025.3624609
BibTeX
@article{Rauscher2026,
author = {Rauscher, Lukas U and Allen, Magdelena S and Judenhofer, Martin S and Byars, Larry and Scipioni, Michele and Udias, José M and Catana, Ciprian and Pichler, Bernd J and Schmidt, Fabian P},
title = {Impact of DOI Capability on Detector Performance: A Comparative Study of DOI and Non-DOI Detectors for High-Resolution and Sensitivity Organ-Specific PET Inserts},
journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
year = {2026},
volume = {10},
number = {5},
pages = {753--763},
doi = {10.1109/trpms.2025.3624609}
}
EUVIMED