Development of a Preclinical PET Scanner With DOI Capabilities Using Semi-Monolithic Detectors

Neus Cucarella, Marta Freire, Antonio J. González, David Sánchez, Andrea Gonzalez-Montoro, Santiago Jiménez-Serrano, Francis Loignon-Houle, Laura Moliner, Celia Valladares, Peter Bruyndonckx, Carlos Correcher, José M. Benlloch

IEEE Transactions on Radiation and Plasma Medical Sciences · 2026-09 · 10 (7) · 1143-1151

DOI: 10.1109/trpms.2025.3648052 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Radiation and Plasma Medical Sciences, 2026-09. Verfasst von Neus Cucarella und 11 weiteren Autor:innen. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: HORIZON EUROPE European Research Council.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Most preclinical PET scanners are based on pixelated detectors without Depth of Interaction (DOI) capabilities, which is crucial to correct for parallax errors. Semi-monolithic crystals have the potential to combine the timing capabilities of pixelated crystals and the 3D positioning accuracy of monolithic scintillators. In this work, we present a preclinical PET prototype consisting of 2 rings defining an inner diameter of 106 mm and an axial length of 52 mm. Each ring contains 14 arrays of 1 × 22 LYSO slabs of 0.97 mm × 25.6 mm × 12 mm, coupled to 8 × 8 SiPMs arrays. The sensitivity, spatial resolution, count rate performance and image quality were studied using the NEMA NU 4-2008 protocol. All images were reconstructed using the MLEM algorithm with 0.5 mm voxel size, including DOI information and normalization correction. A mean spatial resolution for all measured positions and across the three directions (axial, transaxial and radial) of 1.61 ± 0.19 mm was obtained at the center of the FOV. A peak sensitivity of 3.5% was obtained at the center of the scanner, for an energy window (EW) between 358 keV -664 keV. The noise equivalent count rate peak reached 106.9 kcps for an activity of 11.7 MBq using the mouse-sized phantom, the same EW window and a time coincidence window of 10 ns. For the image quality phantom, contrast recovery coefficients of 0.20, 0.62, 0.75, 0.81 and 0.85 were found for the 1, 2, 3, 4 and 5 mm rods, respectively. Spill-over-ratio values of 0.10 for air-filled and 0.23 for water-filled cylinders were measured. Also, according to the Rayleigh criterion, 1.5, 1.2 and 1 mm hot spots of a micro-Derenzo phantom were well distinguished.

Förderung

  • HORIZON EUROPE European Research Council

Zitieren

APA

Cucarella, N., Freire, M., González, A. J., Sánchez, D., Gonzalez-Montoro, A., Jiménez-Serrano, S., Loignon-Houle, F., Moliner, L., Valladares, C., Bruyndonckx, P., Correcher, C., & Benlloch, J. M. (2026). Development of a Preclinical PET Scanner With DOI Capabilities Using Semi-Monolithic Detectors. IEEE Transactions on Radiation and Plasma Medical Sciences, 10(7), 1143-1151. https://doi.org/10.1109/trpms.2025.3648052

Vancouver

Cucarella N, Freire M, González AJ, Sánchez D, Gonzalez-Montoro A, Jiménez-Serrano S, et al. Development of a Preclinical PET Scanner With DOI Capabilities Using Semi-Monolithic Detectors. IEEE Transactions on Radiation and Plasma Medical Sciences. 2026;10(7):1143-1151. doi:10.1109/trpms.2025.3648052

BibTeX
@article{Cucarella2026,
  author = {Cucarella, Neus and Freire, Marta and González, Antonio J. and Sánchez, David and Gonzalez-Montoro, Andrea and Jiménez-Serrano, Santiago and Loignon-Houle, Francis and Moliner, Laura and Valladares, Celia and Bruyndonckx, Peter and Correcher, Carlos and Benlloch, José M.},
  title = {Development of a Preclinical PET Scanner With DOI Capabilities Using Semi-Monolithic Detectors},
  journal = {IEEE Transactions on Radiation and Plasma Medical Sciences},
  year = {2026},
  volume = {10},
  number = {7},
  pages = {1143--1151},
  doi = {10.1109/trpms.2025.3648052}
}
Kostenlos starten →