EUVIMEDEUVIMED entdecken
Open Access

A Low-Noise Low-Power 0.001Hz–1kHz Neural Recording System-on-Chip With Sample-Level Duty-Cycling

Jiajia Wu, Abraham Akinin, Jonathan Somayajulu, Min S Lee, Akshay Paul, Hongyu Lu, Yongjae Park, Seong-Jin Kim, Patrick P Mercier, Gert Cauwenberghs

IEEE Transactions on Biomedical Circuits and Systems · 2024-04 · 18 (2) · 263-273

DOI: 10.1109/tbcas.2024.3368068 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IEEE Transactions on Biomedical Circuits and Systems, 2024-04. Verfasst von Jiajia Wu und 9 weiteren Autor:innen. Hauptthemen laut MEDLINE: Electroencephalography, Brain Waves. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NINDS NIH HHS, NIDA NIH HHS, NIBIB NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Advances in brain-machine interfaces and wearable biomedical sensors for healthcare and human-computer interactions call for precision electrophysiology to resolve a variety of biopotential signals across the body that cover a wide range of frequencies, from the mHz-range electrogastrogram (EGG) to the kHz-range electroneurogram (ENG). Existing integrated wearable solutions for minimally invasive biopotential recordings are limited in detection range and accuracy due to trade-offs in bandwidth, noise, input impedance, and power consumption. This article presents a 16-channel wide-band ultra-low-noise neural recording system-on-chip (SoC) fabricated in 65nm CMOS for chronic use in mobile healthcare settings that spans a bandwidth of 0.001 Hz to 1 kHz through a featured sample-level duty-cycling (SLDC) mode. Each recording channel is implemented by a delta-sigma analog-to-digital converter (ADC) achieving 1.0 μ V rms input-referred noise over 1Hz-1kHz bandwidth with a Noise Efficiency Factor (NEF) of 2.93 in continuous operation mode. In SLDC mode, the power supply is duty-cycled while maintaining consistently low input-referred noise levels at ultra-low frequencies (1.1 μV rms over 0.001Hz-1Hz) and 435 M Ω input impedance. The functionalities of the proposed SoC are validated with two human electrophysiology applications: recording low-amplitude electroencephalogram (EEG) through electrodes fixated on the forehead to monitor brain waves, and ultra-slow-wave electrogastrogram (EGG) through electrodes fixated on the abdomen to monitor digestion.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectroencephalographyBrain WavesHumansEquipment DesignElectrodesElectric ImpedanceAmplifiers, Electronic

Förderung

  • NINDS NIH HHS
  • NIDA NIH HHS
  • NIBIB NIH HHS
  • National Institutes of Health NIMH/NINDS
  • Defense Advanced Research Projects Agency Massive Cross-Correlation
  • Air Force Office of Scientific Research
  • Lawrence Livermore National Laboratory

Zitieren

APA

Wu, J., Akinin, A., Somayajulu, J., Lee, M. S., Paul, A., Lu, H., Park, Y., Kim, S. J., Mercier, P. P., & Cauwenberghs, G. (2024). A Low-Noise Low-Power 0.001Hz–1kHz Neural Recording System-on-Chip With Sample-Level Duty-Cycling. IEEE Transactions on Biomedical Circuits and Systems, 18(2), 263-273. https://doi.org/10.1109/tbcas.2024.3368068

Vancouver

Wu J, Akinin A, Somayajulu J, Lee MS, Paul A, Lu H, et al. A Low-Noise Low-Power 0.001Hz–1kHz Neural Recording System-on-Chip With Sample-Level Duty-Cycling. IEEE Transactions on Biomedical Circuits and Systems. 2024;18(2):263-273. doi:10.1109/tbcas.2024.3368068

BibTeX
@article{Wu2024,
  author = {Wu, Jiajia and Akinin, Abraham and Somayajulu, Jonathan and Lee, Min S and Paul, Akshay and Lu, Hongyu and Park, Yongjae and Kim, Seong-Jin and Mercier, Patrick P and Cauwenberghs, Gert},
  title = {A Low-Noise Low-Power 0.001Hz–1kHz Neural Recording System-on-Chip With Sample-Level Duty-Cycling},
  journal = {IEEE Transactions on Biomedical Circuits and Systems},
  year = {2024},
  volume = {18},
  number = {2},
  pages = {263--273},
  doi = {10.1109/tbcas.2024.3368068}
}