Optimization of Bonner sphere spectrometer design applying Bayesian approach
Radiation Protection Dosimetry · 2025
DOI: 10.1093/rpd/ncaf008 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2025. Verfasst von Ngoc-Quynh Nguyen und 2 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Nuclear Science and Technology, 179 Hoang Quoc Viet, Cau Giay, Hanoi 100000, Viet Nam. Hauptthemen laut MEDLINE: Radiation Monitoring, Neutrons, Americium, Spectrum Analysis. Die Arbeit nennt 15 Literaturangaben. Gefördert von: Vietnam National Foundation for Science and Technology Development.
Abstract (Originalsprache)
The design of Bonner Sphere Spectrometer (BSS) was optimized using Bayesian approach for two different measurement purposes in the ISO 8529 neutron standard field of 241Am-Be source as follows: (i) for the best estimation of neutron fluence spectrum and (ii) for the best estimation of neutron ambient dose equivalent, H*(10). As results, two optimal sets of BSS system (with sequential specific diameters) were proposed: (i) for the first problem, an optimal BSS system with a maximum of six different diameters is needed to obtain neutron fluence spectrum with the standard uncertainties in each energy bin as low as reasonably achievable and (ii) for the second problem, if one would just like to achieve the output H*(10) with a standard uncertainty <10%, only two optimal BSS measurements are needed.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Institute for Nuclear Science and Technology, 179 Hoang Quoc Viet, Cau Giay, Hanoi 100000, Viet Nam.
- National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology, 1-1-1 Umezono, Tsukuba 305-8568, Japan.
ORCID-Profile: Ngoc-Thiem Le
Förderung
- Vietnam National Foundation for Science and Technology Development
Zitieren
APA
Nguyen, N. Q., Kurosawa, T., & Le, N. T. (2025). Optimization of Bonner sphere spectrometer design applying Bayesian approach. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncaf008
Vancouver
Nguyen NQ, Kurosawa T, Le NT. Optimization of Bonner sphere spectrometer design applying Bayesian approach. Radiation Protection Dosimetry. 2025. doi:10.1093/rpd/ncaf008
BibTeX
@article{Nguyen2025,
author = {Nguyen, Ngoc-Quynh and Kurosawa, Tadahiro and Le, Ngoc-Thiem},
title = {Optimization of Bonner sphere spectrometer design applying Bayesian approach},
journal = {Radiation Protection Dosimetry},
year = {2025},
doi = {10.1093/rpd/ncaf008}
}