Exceeding radiation thresholds for cataract induction in diagnostic imaging: a paediatric case report

Amanda Alison Perdomo iD, Mohamed Khaldoun Badawy, Cristina Mignone, Jake Cameron Forster

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncad310 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Amanda Alison Perdomo und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, University of Adelaide, Adelaide 5005, South Australia, Australia. Hauptthemen laut MEDLINE: Cataract, Lens, Crystalline. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This case report investigates the radiation dose received by a paediatric patient with a ventricular assist device who underwent four non-contrast brain computed tomography (CT) scans, two brain perfusion CT scans and two head angiographic CT scans. The total estimated absorbed dose to the lens of the eye is above the 500 mGy radiation-induced cataract threshold. It is recommended that this patient and those with similar imaging histories have routine follow-up with an ophthalmologist. It is also recommended that radiation dose tracking and an electronic medical alert program be implemented to allow the identification of patients who may exceed tissue reaction thresholds.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CataractLens, CrystallineHumansChildRadiation DosageTomography, X-Ray ComputedHead

Autor:innen und Institutionen

  • Department of Physics, University of Adelaide, Adelaide 5005, South Australia, Australia.
  • Department of Medical Imaging, The Royal Children's Hospital, 50 Flemington Road, Parkville 3052, Victoria, Australia.
  • Monash Health Imaging, Monash Health, 246 Clayton Rd, Clayton 3168, Victoria, Australia.
  • Department of Medical Imaging and Radiation Sciences, School of Primary and Allied Health Care, Faculty of Medicine, Nursing and Health Sciences, Monash University, Clayton 3800, VIC, Australia.
  • Medical Physics & Radiation Safety, South Australia Medical Imaging, Adelaide 5005, South Australia, Australia.

ORCID-Profile: Amanda Alison Perdomo

Zitieren

APA

Perdomo, A. A., Badawy, M. K., Mignone, C., & Forster, J. C. (2024). Exceeding radiation thresholds for cataract induction in diagnostic imaging: a paediatric case report. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncad310

Vancouver

Perdomo AA, Badawy MK, Mignone C, Forster JC. Exceeding radiation thresholds for cataract induction in diagnostic imaging: a paediatric case report. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncad310

BibTeX
@article{Perdomo2024,
  author = {Perdomo, Amanda Alison and Badawy, Mohamed Khaldoun and Mignone, Cristina and Forster, Jake Cameron},
  title = {Exceeding radiation thresholds for cataract induction in diagnostic imaging: a paediatric case report},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncad310}
}
Kostenlos starten →