Reduction of radiation dose to the eye lens during common CT examinations of the head, paranasal sinus, and cervical spine in emergency settings: A phantom study
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70486 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Sowitchaya Huakham und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumthani, Thailand. Hauptthemen laut MEDLINE: Lens, Crystalline, Tomography, X-Ray Computed, Phantoms, Imaging, Cervical Vertebrae. Die Arbeit nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Computed tomography (CT) examinations of the head, paranasal sinus (PNS), and cervical spine (C‐spine) are frequently performed in emergency settings, raising concerns about radiation exposure to the radiosensitive eye lens. Overexposure can cause radiation‐induced ocular damage. To address this concern, organ dose modulation (ODM) has emerged as a promising technique for reducing eye lens dose in CT examinations.
Purpose
This study aimed to evaluate radiation exposure to the eye lens and objective image quality metrics for head, PNS, and C‐spine CT examinations using fixed tube current, automatic tube current modulation (ATCM), and ODM techniques.
Methods
Eye lens doses were measured using nanoDot optically stimulated luminescence dosimeters (OSLDs) placed bilaterally to the eye lens of a whole‐body anthropomorphic phantom (PBU‐60). CT scans were performed using a Revolution EX CT scanner with three scanning techniques: fixed tube current, ATCM, and ODM. The phantom was scanned twice for each examination type (head, PNS, and C‐spine) with all three techniques. Eye lens dose reductions with the ODM technique were quantified relative to fixed tube current and ATCM techniques. Image quality was quantitatively evaluated in terms of image noise, signal‐to‐noise ratio (SNR), and contrast‐to‐noise ratio (CNR).
Results
Mean eye lens doses ± standard deviation (SD) using the ODM technique were 38.44 ± 1.37, 17.92 ± 1.01, and 9.77 ± 0.38 mGy for head, PNS, and C‐spine, respectively. These eye lens doses were reduced by 4.28%, 21.33%, and 47.97% compared to the fixed tube current techniques and by 19.40%, 24.70%, and 13.69% compared to the ATCM techniques, for head, PNS, and C‐spine, respectively. These dose reductions were achieved while maintaining image quality metrics (image noise, SNR, and CNR) with no statistically significant differences ( p > 0.05) compared to fixed tube current and ATCM techniques.
Conclusion
Implementation of the ODM technique resulted in significant eye lens dose reduction (4.28%–47.97%) across head, PNS, and C‐spine CT examinations with no significant differences in image noise, SNR, and CNR compared to both fixed tube current and ATCM techniques. ODM demonstrates potential as a practical dose optimization strategy for routine emergency head and neck CT imaging. Further studies with subjective image quality assessment are recommended to evaluate clinical diagnostic acceptability in hospital settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Technology, Faculty of Allied Health Sciences, Thammasat University, Pathumthani, Thailand.
- Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand.
- Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Mueang, Phitsanulok, Thailand.
ORCID-Profile: Sowitchaya Huakham, Thanyawee Pengpan
Zitieren
APA
Huakham, S., Sripoori, W., Suksi, R., Thaikan, T., & Pengpan, T. (2026). Reduction of radiation dose to the eye lens during common CT examinations of the head, paranasal sinus, and cervical spine in emergency settings: A phantom study. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70486
Vancouver
Huakham S, Sripoori W, Suksi R, Thaikan T, Pengpan T. Reduction of radiation dose to the eye lens during common CT examinations of the head, paranasal sinus, and cervical spine in emergency settings: A phantom study. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70486
BibTeX
@article{Huakham2026,
author = {Huakham, Sowitchaya and Sripoori, Wirachad and Suksi, Raksumon and Thaikan, Thawatchai and Pengpan, Thanyawee},
title = {Reduction of radiation dose to the eye lens during common CT examinations of the head, paranasal sinus, and cervical spine in emergency settings: A phantom study},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70486}
}
EUVIMED