EUVIMEDEUVIMED entdecken
Crossref

Modification of algorithm for accurate estimation of H p(3) based on response of head TLD badge calibrated using ISO cylindrical phantom

Madhumita Bhattacharya iD, Kasthuri Samuel, Kshama Srivastava, Ashok Kumar Bakshi

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncae164 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Madhumita Bhattacharya und 3 weiteren Autor:innen. Beteiligte Einrichtung: Radiological Physics and Advisory Division, Bhabha Atomic and Research Centre (BARC), Mumbai 400085, India. Hauptthemen laut MEDLINE: Algorithms, Phantoms, Imaging, Lens, Crystalline, Thermoluminescent Dosimetry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In India, the prevailing approach to eye lens dosimetry is the placement of an existing dosemeter on the forehead region after slight modification serves as a dedicated Eye Lens Dosemeter. A methodology for estimating the eye lens dose in terms of the Hp(3) has been previously explored employing an algorithm based on the response characteristics of this dosemeter using ISO slab phantom. It was observed that the performance of the dosemeter in terms of Hp(3) using previous algorithm showed under response at higher angles of incidence and photon beams of energy < 200 keV. Further, study was conducted to modify the algorithm following the latest ISO recommendations. This involved generation of data from the response of existing dosemeter on a cylindrical phantom. The results of this study revealed better performance of the newly established algorithm in estimating eye lens dose in terms of Hp(3) when compared to the earlier algorithm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsPhantoms, ImagingLens, CrystallineThermoluminescent DosimetryHumansCalibrationRadiation DosagePhotonsHeadIndiaRadiation Protection

Autor:innen und Institutionen

  • Radiological Physics and Advisory Division, Bhabha Atomic and Research Centre (BARC), Mumbai 400085, India.
  • Homi Bhabha National Institute, Anushaktinagar, Mumbai 400094, India.

ORCID-Profile: Madhumita Bhattacharya

Zitieren

APA

Bhattacharya, M., Samuel, K., Srivastava, K., & Bakshi, A. K. (2024). Modification of algorithm for accurate estimation of H p(3) based on response of head TLD badge calibrated using ISO cylindrical phantom. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncae164

Vancouver

Bhattacharya M, Samuel K, Srivastava K, Bakshi AK. Modification of algorithm for accurate estimation of H p(3) based on response of head TLD badge calibrated using ISO cylindrical phantom. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncae164

BibTeX
@article{Bhattacharya2024,
  author = {Bhattacharya, Madhumita and Samuel, Kasthuri and Srivastava, Kshama and Bakshi, Ashok Kumar},
  title = {Modification of algorithm for accurate estimation of H p(3) based on response of head TLD badge calibrated using ISO cylindrical phantom},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncae164}
}
Kostenlos starten →