Effectiveness of low tube voltage scan in the exposure dose for lenses during paediatric thoracic CT examination: anthropomorphic phantoms study

Takanori Masuda iD, Masao Kiguchi, Chikako Fujioka, Takayuki Oku iD, Toru Ishibashi iD, Yasushi Katsunuma, Takayasu Yoshitake, Shuji Abe, Kazuo Awai

Radiation Protection Dosimetry · 2024

DOI: 10.1093/rpd/ncad284 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Radiation Protection Dosimetry, 2024. Verfasst von Takanori Masuda und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, 288, Matsushima, Kurashiki, …. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Lens, Crystalline. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Japanese Society of Radiological Technology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To determine whether using lower-tube voltage reduces the scattered dose for the lens during paediatric thoracic computed tomography (CT). Two paediatric anthropomorphic phantoms (ATOM Phantom, CIRS, Norfolk, Virginia, USA) representing a newborn and 5-year-old were placed on the gantry of CT scanner, and optically stimulated luminescence dosemeters were placed on the left and right lenses, in front of the left and right thyroid glands, in front of the left and right mammary glands, and in front of and behind the mammary gland level and we measured scattered dose of the optically stimulated luminescence dosemeter was compared for each phantom between 80 and 120 kVp. Significant differences were observed in the scatter doses for the lens between 80 and 120 kVp (p < 0.01). Compared with the 120 kVp scan, the scatter doses for the lens were ~15–40% lower in newborn and 5-year-olds using the 80 kVp scan during paediatric CT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedLens, CrystallineInfant, NewbornChildHumansChild, PreschoolRadiation DosagePhantoms, ImagingRadionuclide Imaging

Autor:innen und Institutionen

  • Department of Radiological Technology, Faculty of Health Science and Technology, Kawasaki University of Medical Welfare, 288, Matsushima, Kurashiki, Okayama 701-0193, Japan.
  • Department of Radiology, Hiroshima University, 2-3, Kasumi, Minami-ku, Hiroshima 734-0037, Japan.
  • Department of Radiological Technologist, Tsuchiya General Hospital, Nakajima-cho 3-30, Naka-ku, Hiroshima 730-8655, Japan.
  • Department of Radiological Technology, Tokai University Oiso Hospital, 143, Shimokasuya, Iseharashi, Kanagawa 259-1193, Japan.
  • TOWCAR WORKS Co., Ltd Hoashi 233-1, Kusu-machi, Kusugun, Oita 879-4403, Japan.
  • Department of Radiological Technologist, Osaka College of High Technology, 1-2-43, Miyahara, Yogogawa-ku, Osaka 532-0003, Japan.

ORCID-Profile: Takanori Masuda, Takayuki Oku, Toru Ishibashi

Förderung

  • Japanese Society of Radiological Technology

Zitieren

APA

Masuda, T., Kiguchi, M., Fujioka, C., Oku, T., Ishibashi, T., Katsunuma, Y., Yoshitake, T., Abe, S., & Awai, K. (2024). Effectiveness of low tube voltage scan in the exposure dose for lenses during paediatric thoracic CT examination: anthropomorphic phantoms study. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncad284

Vancouver

Masuda T, Kiguchi M, Fujioka C, Oku T, Ishibashi T, Katsunuma Y, et al. Effectiveness of low tube voltage scan in the exposure dose for lenses during paediatric thoracic CT examination: anthropomorphic phantoms study. Radiation Protection Dosimetry. 2024. doi:10.1093/rpd/ncad284

BibTeX
@article{Masuda2024,
  author = {Masuda, Takanori and Kiguchi, Masao and Fujioka, Chikako and Oku, Takayuki and Ishibashi, Toru and Katsunuma, Yasushi and Yoshitake, Takayasu and Abe, Shuji and Awai, Kazuo},
  title = {Effectiveness of low tube voltage scan in the exposure dose for lenses during paediatric thoracic CT examination: anthropomorphic phantoms study},
  journal = {Radiation Protection Dosimetry},
  year = {2024},
  doi = {10.1093/rpd/ncad284}
}
Kostenlos starten →