Investigation of the effect of thyroid collar, radiation safety glasses, and lead apron on radiation dose in cone beam CT

Derya İçöz iD, Osman Vefa Gül iD

Dentomaxillofacial Radiology · 2025

DOI: 10.1093/dmfr/twaf007 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Dentomaxillofacial Radiology, 2025. Verfasst von Derya İçöz und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Dentistry, Department of Oral and Maxillofacial Radiology, Selcuk University, Konya 42130, Turkey. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Thyroid Gland, Radiation Dosage, Radiation Protection. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 30 Literaturangaben. Gefördert von: Selcuk University Scientific Research Projects Coordination.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

Due to the increasing use of cone-beam CT (CBCT) in dentistry and considering the effects of radiation on radiosensitive organs, the aim of this study was to investigate the effect of shielding on absorbed dose of eyes, thyroid, and breasts in scans conducted with different parameters using 2 different fields of view (FOV).

Methods

Dose measurements were calculated on a tissue-equivalent female phantom by repeating each scanning parameter 3 times and placing at least 2 thermoluminescent dosimeters (TLD) on each organ, with the averages then taken. The same CBCT scans were performed in 2 different FOV with shielding including thyroid collar, radiation safety glasses, and lead apron and without shielding. The differences between them were analysed statistically. Descriptive statistics and the Wilcoxon test were used for data analysis.

Results

The difference between measurements with and without shielding was statistically significant for all scans (P < .001). The dose reduction associated with the use of shielding ranged from 26.81% to 52.95%. The dose related to the FOV has shown a significant increase, ranging from 8.30% to 623.54%, due to both the variation in the area affected by the primary beam on the organs and changes in the amount of radiation.

Conclusion

There are significant differences in the absorbed dose depending on shielding and FOV usage. Therefore, the CBCT imaging protocol should be optimized by the operator, and special attention should be paid to the use of thyroid collars and radiation safety glasses, considering their effects on image quality.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyThyroid GlandRadiation DosageRadiation ProtectionHumansPhantoms, ImagingFemaleThermoluminescent DosimetryBreastLeadEye

Stichwörter der Autor:innen

cone-beam CTradiation doseradiation dosimetryradiation protection

Substanzen

Lead

Autor:innen und Institutionen

  • Faculty of Dentistry, Department of Oral and Maxillofacial Radiology, Selcuk University, Konya 42130, Turkey.
  • Faculty of Medicine, Department of Radiation Oncology, Selcuk University, Konya 42130, Turkey.

ORCID-Profile: Derya İçöz, Osman Vefa Gül

Förderung

  • Selcuk University Scientific Research Projects Coordination

Zitieren

APA

İçöz, D., & Gül, O. V. (2025). Investigation of the effect of thyroid collar, radiation safety glasses, and lead apron on radiation dose in cone beam CT. Dentomaxillofacial Radiology. https://doi.org/10.1093/dmfr/twaf007

Vancouver

İçöz D, Gül OV. Investigation of the effect of thyroid collar, radiation safety glasses, and lead apron on radiation dose in cone beam CT. Dentomaxillofacial Radiology. 2025. doi:10.1093/dmfr/twaf007

BibTeX
@article{z2025,
  author = {İçöz, Derya and Gül, Osman Vefa},
  title = {Investigation of the effect of thyroid collar, radiation safety glasses, and lead apron on radiation dose in cone beam CT},
  journal = {Dentomaxillofacial Radiology},
  year = {2025},
  doi = {10.1093/dmfr/twaf007}
}
Kostenlos starten →