A survey and analysis of radiotherapy linear accelerator quality control in the United Kingdom, for c-arm and ring gantry systems and associated patient specific quality control
British Journal of Radiology · 2025
DOI: 10.1093/bjr/tqaf140 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Radiology, 2025. Verfasst von Elisha Tassano-Smith und 3 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire PO63LY, United Kingdom. Hauptthemen laut MEDLINE: Particle Accelerators, Quality Control, Radiotherapy. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Objectives
To conduct a survey of radiotherapy linear accelerator quality control (QC) across the United Kingdom (UK) on behalf of the Institute of Physics and Engineering in Medicine (IPEM) Radiotherapy Special Interest Group and Interdepartmental Dosimetry Audit (IDA) Sub-committee. To update results from a similar survey published in 2012 and compare to the latest guidance from IPEM Report 81 (2018). There have been significant developments of equipment and clinical practice since the previous survey and IPEM publication, requiring an updated review and benchmark of QC practice.
Methods
All UK radiotherapy centres were invited to complete a comprehensive survey of their local QC practice, with questions on c-arm gantry, ring-gantry, linac ancillary equipment, and patient-specific QC.
Results
63% (n = 43/68) of the UK radiotherapy centres responded. IPEM Report 81 was used to inform QC practice in 91% of centres. For the majority of tests studied centres were meeting or exceeding the recommendations of this report. Standard output was still performed weekly in 26% of centres compared to monthly recommendation in Report 81. Comprehensive tables of frequency and tolerances of QC tests were collated for c-arm and ring gantry linacs and ancillary equipment.
Conclusions
A comprehensive review of consensus practice for linac QC radiotherapy across the UK is presented. Findings include the main stated reasons QC is undertaken is to “demonstrate safe use.” On efficiency, it was found that about half of centres state they undertake “the right amount of QC.” Half also state review of their QC process is “required.” Advances in knowledge Updated data are presented on current practice for linac QC in the UK.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire PO63LY, United Kingdom.
- Department of Medical Physics and Biomedical Engineering, University College London, London WC1E6BT, United Kingdom.
ORCID-Profile: Elisha Tassano-Smith, Antony L Palmer, Wojciech Polak, Andrew Nisbet
Zitieren
APA
Tassano-Smith, E., Palmer, A. L., Polak, W., & Nisbet, A. (2025). A survey and analysis of radiotherapy linear accelerator quality control in the United Kingdom, for c-arm and ring gantry systems and associated patient specific quality control. British Journal of Radiology. https://doi.org/10.1093/bjr/tqaf140
Vancouver
Tassano-Smith E, Palmer AL, Polak W, Nisbet A. A survey and analysis of radiotherapy linear accelerator quality control in the United Kingdom, for c-arm and ring gantry systems and associated patient specific quality control. British Journal of Radiology. 2025. doi:10.1093/bjr/tqaf140
BibTeX
@article{TassanoSmith2025,
author = {Tassano-Smith, Elisha and Palmer, Antony L and Polak, Wojciech and Nisbet, Andrew},
title = {A survey and analysis of radiotherapy linear accelerator quality control in the United Kingdom, for c-arm and ring gantry systems and associated patient specific quality control},
journal = {British Journal of Radiology},
year = {2025},
doi = {10.1093/bjr/tqaf140}
}
EUVIMED