A review of kilovoltage radiotherapy treatment in the United Kingdom: quality control, radiation dosimetry, treatment equipment, and workload

Antony L Palmer iD, Jason Brimelow iD, Patrick Downes iD, Mayur Munshi, David Nash iD, Bhupinder Rai, Andrew R White iD

British Journal of Radiology · 2025

DOI: 10.1093/bjr/tqaf001 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift British Journal of Radiology, 2025. Verfasst von Antony L Palmer und 6 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire, PO63LY, United Kingdom. Hauptthemen laut MEDLINE: Radiometry, Radiotherapy, Workload. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objectives

To survey kilovoltage (kV) radiotherapy in the United Kingdom, updating a 2016 study, focussing on radiotherapy physics, including equipment quality control (QC) and radiation dosimetry, with information on installed equipment and clinical activity.

Methods

All UK radiotherapy physics departments (n = 68) were invited to complete a comprehensive survey. An analysis of the installed equipment base, patient numbers, clinical activity, QC testing, and radiation dosimetry processes were undertaken.

Results

91% of centres (n = 62) responded to the survey. kV radiotherapy was available in 70% of UK radiotherapy departments, with a wide variation in workload; 7-436 patients/centre annually. There has been an increase in centres using treatment calculation software rather than manual methods, up from 36% in 2016 to 50% currently. Only 50% of centres use an independent calculation check method. There was an increase in the use of the addendum to the UK dosimetry code of practice, enabling medium energy calibration in-air rather than at depth in phantom, citing “clinical relevance.” Appropriate levels of QC testing were being conducted at UK centres, with Institute of Physics and Engineering in Medicine (IPEM) Report 81 cited as a primary source of guidance. Good consensus for the frequency and tolerance values used for QC was seen across UK centres.

Conclusions

A comprehensive review of consensus practice for QC and dosimetry in kV radiotherapy across the United Kingdom is presented, with supporting information on equipment installation and clinical use. Advances in Knowledge Updated data are presented on kV radiotherapy treatment in the United Kingdom, with focus on physics aspects of QC and dosimetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiometryRadiotherapyWorkloadUnited KingdomHumansQuality ControlRadiotherapy DosageSurveys and Questionnaires

Stichwörter der Autor:innen

United KingdomkV radiotherapyquality controlradiation dosimetrysurvey

Autor:innen und Institutionen

  • Medical Physics Department, Portsmouth Hospitals University NHS Trust, Queen Alexandra Hospital, Portsmouth, Hampshire, PO63LY, United Kingdom.
  • Department of Medical Physics and Biomedical Engineering, University College London, London, WC1E6BT, United Kingdom.
  • Medical Physics Department, Cheltenham General Hospital, Cheltenham, Gloucestershire, GL537AN, United Kingdom.
  • Radiotherapy Physics, Northern Centre for Cancer Care, Freeman Hospital, Newcastle upon Tyne, Northumberland, NE77DN, United Kingdom.
  • Radiotherapy Physics, Department of Clinical Physics and Bio-engineering, Beatson West of Scotland Cancer Centre, Glasgow, Lanarkshire, G120YN, United Kingdom.
  • Radiotherapy Physics Department, University College London Hospital, London, WC1E6BT, United Kingdom.

ORCID-Profile: Antony L Palmer, Jason Brimelow, Patrick Downes, David Nash, Andrew R White

Zitieren

APA

Palmer, A. L., Brimelow, J., Downes, P., Munshi, M., Nash, D., Rai, B., & White, A. R. (2025). A review of kilovoltage radiotherapy treatment in the United Kingdom: quality control, radiation dosimetry, treatment equipment, and workload. British Journal of Radiology. https://doi.org/10.1093/bjr/tqaf001

Vancouver

Palmer AL, Brimelow J, Downes P, Munshi M, Nash D, Rai B, et al. A review of kilovoltage radiotherapy treatment in the United Kingdom: quality control, radiation dosimetry, treatment equipment, and workload. British Journal of Radiology. 2025. doi:10.1093/bjr/tqaf001

BibTeX
@article{Palmer2025,
  author = {Palmer, Antony L and Brimelow, Jason and Downes, Patrick and Munshi, Mayur and Nash, David and Rai, Bhupinder and White, Andrew R},
  title = {A review of kilovoltage radiotherapy treatment in the United Kingdom: quality control, radiation dosimetry, treatment equipment, and workload},
  journal = {British Journal of Radiology},
  year = {2025},
  doi = {10.1093/bjr/tqaf001}
}
Kostenlos starten →