Evaluation of a novel patient-specific quality assurance phantom for robotic single-isocentre, multiple-target stereotactic radiosurgery, and stereotactic radiotherapy
British Journal of Radiology · 2024
DOI: 10.1093/bjr/tqae011 ↗Kurz gefasst
Fachartikel in der Zeitschrift British Journal of Radiology, 2024. Verfasst von Qianyi Xu und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Advanced Radiation Oncology and Proton Therapy, Inova Schar Cancer Institute, Fairfax, VA 22031, United States. Hauptthemen laut MEDLINE: Radiosurgery, Robotic Surgical Procedures, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objectives
To evaluate patient-specific quality assurance (PSQA) of 3 targets in a single delivery using a novel film-based phantom.
Methods
The phantom was designed to rotate freely as a sphere and could measure 3 targets with film in a single delivery. After identifying the coordinates of 3 targets in the skull, the rotation angles about the equator and meridian were computed for optimal phantom setup, ensuring the film plane intersected the 3 targets. The plans were delivered on the CyberKnife system using fiducial tracking. The irradiated films were scanned and processed. All films were analysed using 3 gamma criteria.
Results
Fifteen CyberKnife test plans with 3 different modalities were delivered on the phantom. Both automatic and marker-based registration methods were applied when registering the irradiated film and dose plane. Gamma analysis was performed using a 3%/1 mm, 2%/1 mm, and 1%/1 mm criteria with a 10% threshold. For the automatic registration method, the passing rates were 98.2% ± 1.9%, 94.2% ± 3.7%, and 80.9% ± 6.3%, respectively. For the marker-based registration approach, the passing rates were 96.4% ± 2.7%, 91.7% ± 4.3%, and 78.4% ± 6.2%, respectively.
Conclusions
A novel spherical phantom was evaluated for the CyberKnife system and achieved acceptable PSQA passing rates using TG218 recommendations. The phantom can measure true-composite dose and offers high-resolution results for PSQA, making it a valuable device for robotic radiosurgery. Advances in knowledge This is the first study on PSQA of 3 targets concurrently on the CyberKnife system.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Advanced Radiation Oncology and Proton Therapy, Inova Schar Cancer Institute, Fairfax, VA 22031, United States.
- Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, PA 19107, United States.
- Department of Radiation Oncology, Wellstar Health System, Marietta, GA 30060, United States.
- Department of Radiation Oncology, Taizhou Hospital, Zhejiang 317000, China.
ORCID-Profile: Qianyi Xu
Zitieren
APA
Xu, Q., Vinogradskiy, Y., Grimm, J., Nie, W., Dupre, P., Chawla, A. K., Bajaj, G., Yang, H., LaCouture, T., & Fan, J. (2024). Evaluation of a novel patient-specific quality assurance phantom for robotic single-isocentre, multiple-target stereotactic radiosurgery, and stereotactic radiotherapy. British Journal of Radiology. https://doi.org/10.1093/bjr/tqae011
Vancouver
Xu Q, Vinogradskiy Y, Grimm J, Nie W, Dupre P, Chawla AK, et al. Evaluation of a novel patient-specific quality assurance phantom for robotic single-isocentre, multiple-target stereotactic radiosurgery, and stereotactic radiotherapy. British Journal of Radiology. 2024. doi:10.1093/bjr/tqae011