Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Canyang Lin und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oncology, Longyan First Affiliated Hospital of Fujian Medical University, Longyan, Fujian, China. Hauptthemen laut MEDLINE: Radiosurgery, Brain Neoplasms, Radiotherapy Planning, Computer-Assisted, Robotic Surgical Procedures. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To compare dosimetric characteristics and treatment efficiency between the high‐definition multileaf collimator (HD‐MLC) system and robotic radiosurgery (RRS) for brain metastases, with stratified analysis by lesion count.
Methods
Twenty‐nine patients (12 single [SBM], 17 multiple [MBM] brain metastases) treated with RRS (CyberKnife) were re‐planned using a linear accelerator (LINAC) with 2.5‐mm micro‐MLC (D2SRS system). Key indices included target parameters (RTOG conformity index [RTOG CI], Paddick conformity index [Paddick CI], homogeneity index [HI], gradient index [GI], dose at 2 cm outside the target [D2cm]) and normal brain doses (V12Gy, V20Gy, V30Gy). The Wilcoxon signed‐rank test was used ( p < 0.05).
Results
In SBM, RTOG CI, Paddick CI, and GI showed no significant differences, but Paddick CI favored RRS for tumor volume (TV) < 80 cc and D2SRS for TV > 80 cc. D2SRS had better HI (0.08 ± 0.02 vs. 0.12 ± 0.03, p < 0.01); RRS had lower D2cm (18.2 ± 5.5 Gy vs. 23.8 ± 6.7 Gy, p = 0.008), V12Gy (5.3 ± 2.1 cc vs. 8.7 ± 3.5 cc, p < 0.01), and V20Gy (18.47 ± 21.17 cc vs. 35.30 ± 53.85 cc, p = 0.015), with similar V30Gy. D2SRS had shorter treatment time (TT) (8.2 ± 2.1 min vs. 26.8 ± 7.8 min, p < 0.01). In MBM, D2SRS outperformed RRS in RTOG CI (1.12 ± 0.21 vs. 1.45 ± 0.33, p = 0.007), Paddick CI (0.85 ± 0.11 vs. 0.68 ± 0.14, p = 0.007), and HI (0.07 ± 0.02 vs. 0.13 ± 0.03, p = 0.03). RRS had lower V12Gy (38.5 ± 12.6 cc vs. 48.1 ± 15.3 cc, p = 0.017), V20Gy (14.5 ± 4.2 cc vs. 21.3 ± 6.8 cc, p = 0.04), and V30Gy (14.99 ± 20.81 cc vs. 18.49 ± 23.19 cc, p = 0.026). D2SRS reduced TT by 77.3% (12.4 ± 3.5 min vs. 54.6 ± 15.2 min, p < 0.01). No significant differences in Dmax of the brainstem, optic nerves, or lenses were found.
Conclusion
RRS provides better normal brain protection in both groups, while D2SRS shows superior conformity and efficiency in MBM. Clinical selection should depend on lesion burden.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oncology, Longyan First Affiliated Hospital of Fujian Medical University, Longyan, Fujian, China.
- Department of Radiation Oncology, Peking University Third Hospital, Beijing, China.
ORCID-Profile: Canyang Lin
Zitieren
APA
Lin, C., Liu, W., Xiao, N., Jiang, Y., & Wang, J. (2025). Stratified dosimetric comparison between high‐definition multileaf collimator‐based D2SRS and robotic radiosurgery in the treatment of brain metastases. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70361
Vancouver
Lin C, Liu W, Xiao N, Jiang Y, Wang J. Stratified dosimetric comparison between high‐definition multileaf collimator‐based D2SRS and robotic radiosurgery in the treatment of brain metastases. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70361
BibTeX
@article{Lin2025,
author = {Lin, Canyang and Liu, Wei and Xiao, Nan and Jiang, Yuliang and Wang, Junjie},
title = {Stratified dosimetric comparison between high‐definition multileaf collimator‐based D2SRS and robotic radiosurgery in the treatment of brain metastases},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70361}
}
EUVIMED