Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT
Journal of Applied Clinical Medical Physics · 2023-12 · 24 (12)
DOI: 10.1002/acm2.14143 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 22.09.2023. Verfasst von Katerine Viviana Díaz Hernández und 2 weiteren Autor:innen. Beteiligte Einrichtung: Science Faculty, University of Zürich, Zürich, Switzerland. Hauptthemen laut MEDLINE: Radiosurgery, Robotic Surgical Procedures, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
The purpose of this study is to assess the quality of automatic planned O‐Ring Halcyon linac SBRT plans for pelvic lymph node metastases and to establish an absolute PTV volume threshold as a plan quality prediction criterion. Compliance of the plans to institutional SBRT plan evaluation criteria and differences in plan quality and treatment delivery times between Halcyon Linac and CyberKnife robotic SBRT were evaluated.
Methods
Twenty‐one CyberKnife treatment plans were replanned for Halcyon. Prescription doses range was 26–40 Gy in mean three fractions. The mean/median planning target volume was 4.0/3.6 cm 3 . Institutional criteria for the plan evaluation were: New Conformity Index (NCI), Conformity Index (CI), Modified Gradient Index (MGI), selectivity index reciprocal (PIV/TV PIV ), and the target coverage by prescription isodose (%PIV). Statistical analysis based on the receiver operating characteristic (ROC) curve was used to determine a plan quality predictor threshold of the PTV volume. Comparative analysis of normal tissue complication probabilities (NTCP) was used to assess the risk of toxicity in healthy tissues.
Results
Seventy‐one percent ( n = 15)/95% ( n = 20) of Halcyon and 81% ( n = 17)/100% ( n = 21) of CK plans fulfilled all ideal/tolerance criteria. For PTVs above a found optimal threshold of 2.6 cm 3 (71%, n = 15), no statistically significant difference was observed between the CI, NCI, PIV/TV PIV , and MGI indexes of both groups, while the coverage (%PIV) was statistically but not clinically significantly different between cohorts. Significantly shorter delivery times are expected with Halcyon. No significant differences in NTCP were observed.
Conclusion
All but one automatically optimized Halcyon treatment plans demonstrated ideal or acceptable performance. PTV threshold of 2.6 cm 3 can be used as decision criteria in clinical settings. The results of our study demonstrated the promising performance of the Halcyon for pelvic SBRT, although plan‐specific QA is required to verify machine performance during plan delivery.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Science Faculty, University of Zürich, Zürich, Switzerland.
- Medical Physics, Radiotherapy Hirslanden, Zürich, Switzerland.
Zitieren
APA
Hernández, K. V. D., Unterkirhers, S., & Schneider, U. (2023). Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT. Journal of Applied Clinical Medical Physics, 24(12). https://doi.org/10.1002/acm2.14143
Vancouver
Hernández KVD, Unterkirhers S, Schneider U. Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT. Journal of Applied Clinical Medical Physics. 2023;24(12). doi:10.1002/acm2.14143
BibTeX
@article{Hernndez2023,
author = {Hernández, Katerine Viviana Díaz and Unterkirhers, Sergejs and Schneider, Uwe},
title = {Quality assessment of automatically planned O‐Ring linac SBRT plans for pelvic lymph node metastases, finding the optimal minimum target size by comparison with robotic SBRT},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
volume = {24},
number = {12},
doi = {10.1002/acm2.14143}
}
EUVIMED