Investigation of organs dosimetry precision using ATOM phantom and optically stimulated luminescence detectors in computed tomography

Abdellah Khallouqi iD, Hamza Sekkat iD, Omar El Rhazouani, Abdellah Halimi

Biomedical Physics & Engineering Express · 2025

DOI: 10.1088/2057-1976/ad992e ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Physics & Engineering Express, 2025. Verfasst von Abdellah Khallouqi und 3 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Health Sciences and Technologies, Higher Institute of Health Sciences, Hassan 1st University, Settat, Morocco. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Radiation Dosage, Optically Stimulated Luminescence Dosimetry. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The primary objective of this study was to compare organ doses measured using optically stimulated luminescent dosimeters (OSLDs) with those estimated by the CT-EXPO software for common CT protocols. An anthropomorphic ATOM phantom was employed to measure organ doses across head, chest, and abdominal CT scans performed on a Hitachi Supria 16-slice CT scanner. These OSLD measurements were then compared to the estimates provided by the widely used CT-EXPO software. Organ doses were assessed using OSLDs placed in an adult anthropomorphic phantom, with calibration performed through a comprehensive process involving multiple tube potentials and sensitivity corrections. Results from three CT acquisitions per protocol were compared to estimates provided by CT-EXPO software. Findings reveal significant discrepancies between measured and estimated organ doses, with p-values consistently below 0.05 across all organs. For head CT, measured eye lens doses averaged 33.51 mGy, 6.0% lower than the estimated 35.65 mGy. In chest CT, the thyroid dose was 9.82 mGy, 13.5% higher than the estimated 8.65 mGy. For abdominal CT, the liver dose measured 12.11 mGy, 9.6% higher than the estimated 11.05 mGy. Measured doses for the rest of organs were generally lower than those predicted by CT-EXPO, showing some limitations in current estimation models and the importance of precise dosimetry. This study highlights the potential of OSLD measurements as a complementary method for organ dose assessment in CT imaging, emphasizing the need for more accurate organ dose measurement to optimize patient care.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingTomography, X-Ray ComputedRadiation DosageOptically Stimulated Luminescence DosimetrySoftwareHumansCalibrationRadiometryHeadAdultLiverThyroid GlandLuminescence

Stichwörter der Autor:innen

OSLDcomputed tomographydosimetryorganphantom

Autor:innen und Institutionen

  • Laboratory of Health Sciences and Technologies, Higher Institute of Health Sciences, Hassan 1st University, Settat, Morocco.

ORCID-Profile: Abdellah Khallouqi, Hamza Sekkat

Zitieren

APA

Khallouqi, A., Sekkat, H., El Rhazouani, O., & Halimi, A. (2025). Investigation of organs dosimetry precision using ATOM phantom and optically stimulated luminescence detectors in computed tomography. Biomedical Physics & Engineering Express. https://doi.org/10.1088/2057-1976/ad992e

Vancouver

Khallouqi A, Sekkat H, El Rhazouani O, Halimi A. Investigation of organs dosimetry precision using ATOM phantom and optically stimulated luminescence detectors in computed tomography. Biomedical Physics & Engineering Express. 2025. doi:10.1088/2057-1976/ad992e

BibTeX
@article{Khallouqi2025,
  author = {Khallouqi, Abdellah and Sekkat, Hamza and El Rhazouani, Omar and Halimi, Abdellah},
  title = {Investigation of organs dosimetry precision using ATOM phantom and optically stimulated luminescence detectors in computed tomography},
  journal = {Biomedical Physics & Engineering Express},
  year = {2025},
  doi = {10.1088/2057-1976/ad992e}
}
Kostenlos starten →